Checking the patient table
Makes sure the following systems operate correctly: patient table up and down, cradle in and out, emergency egress, and touch-and-go strip.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | - |
About this task
- Checking the table up and down functions
- Checking the cradle in and out functions
- Checking the emergency egress function
- Checking the touch-and-go strip function
- Doing the cradle motion test
- Doing a TPS Reset
The table must be docked prior to beginning the checks.