• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30D90450GYZ
  • Topic ID: id_2021618 Version: 1.3
  • Date: Apr 9, 2020 12:11:52 PM

Installing the dockable table dummy load

Installs the dummy load into the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Install the dummy load brackets onto the dummy load.
    Figure 1. Dummy load brackets
    1Brackets
    2Dummy load
  2. Install the dummy load into the dockable table with the BNC connector facing the handle-end of the table.
    Note: The bumper may be stretched slightly to provide better access.
  3. Install the four screws securing the dummy load to the table top.
    Figure 2. Dummy load screws
  4. Connect the BNC connector to the dummy load.
    Figure 3. Dummy load connector
    1Connector
    2Dummy load
  5. Install the three screws securing the bumper to the table.
    Figure 4. Bumper screws
  6. Release the table brakes.
  7. Dock the table.