• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30CA4450GYZ
  • Topic ID: id_2022084 Version: 1.4
  • Date: Dec 20, 2019 11:49:09 AM

Finalizing egress switch and springs replacement

Finalization steps after replacing the egress switch and springs.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1--15 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set1 Kit5113258-

Procedure

  1. Dock the table both mechanically and electrically.
  2. Move the cradle toward the bore a minimum of 750 mm.
  3. Activate the egress switch so that the cradle comes out of the bore, initially at a high speed, and then ramps down to the regular home speed.
    If the egress switch does not activate or does not move the cradle out from the bore, repeat the replacement procedure.