Adjusts the height of the secondary latch to within tolerance.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
10 minutes
-
Tools and test equipment
Item
Quantity
Part number
Manufacturer
Stainless Steel Ruler
1
46-198452P2
-
Nonmagnetic Titanium Service Tool Kit, Small Set
1
5113258
-
Procedure
Dock the table.
Push the cradle into the bore until the secondary latch is exposed.
Figure 1. Secondary latch
1
Secondary latch pin
2
Secondary latch housing
Use the threaded cable adjuster to adjust the cable such that the top surface of the secondary latch pin is 0 to 3 mm below the surface of the latch housing.
Note: A measurement of 0 - 3 mm is within specification for the secondary latch pin. However, a value of 0 - 1 mm must be entered in ICW for this measurement.
Figure 2. Threaded cable adjuster
1
Lock nut
2
Push rod
3
Set screw
Undock the table.
Press and hold the Dock pedal to override the primary latch and push the cradle approximately 80 mm off the tabletop surface.
Push and hold the rod button and pull the cradle off the tabletop surface until the cradle is clear of the secondary latch.
CAUTION
Pinching hazard
If the table moves down unexpectedly, injury could result or the tabletop may be damaged.
Push the button shown in the illustration and not the button on the opposite side of the table.
Figure 3. Cradle rod button
Make sure the latch pin is a minimum of 6.35 mm above the latch housing.
Adjust the cable until the docked and undocked measurements are both within limits.