• Object ID: 00000018WHA307372GYZ
  • Topic ID: id_2001557 Version: 2.2
  • Date: Jan 7, 2021 3:19:53 AM

Finalizing the Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) replacement

This procedure finalizes the Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) replacement.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1---

Procedure

  1. Restore the power. Refer to the MR Service Safety Manual (5452735).
  2. Do the following Receiver Diagnostics and confirm that all of the tests pass.
    • Receive Chain BLD
    • Receive Chain DP
    • TNS DP
    • ICE to RRx+DTX DP
    • RF Calibration
  3. Do DQA II Tool.
  4. Do a Check Scan.