Removing the dockable table Physiological Acquisition Controller (PAC)
Removes the Physiological Acquisition Controller (PAC) from the dockable table.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
15 minutes
-
Tools and test equipment
Item
Quantity
Part number
Manufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set
1
5113258
-
Required conditions
The right-hand base side cover has been removed.
The table is disconnected or ISC LOTO is applied.
Procedure
If it is not already undocked, undock the table and move it away from the magnet.
Important: Undocking the table disconnects the power to the Physiological Acquisition Controller (PAC). If the table is docked during this procedure, ISC LOTO is required.
Disconnect the following cables and pneumatic tube from the PAC:
Figure 1. PAC connections
1
ECG extension cable
PAC ECG connector
2
M3386
J6
3
M3386
J5
4
5946424
J2
5
Resp tube
J3
6
5946423
J1
7
5946424
J9
Remove the four screws and four washers securing the PAC to the table.
Figure 2. PAC hardware
1
Washer
2
Screw
Cut cable ties securing cables to the PAC bracket.