- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA303E1030GYZ
- id_131069161.7
- Jul 5, 2019 10:46:04 PM
Slice Thickness and Resolution
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 30 minutes | Not Applicable |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Axial Slice Analysis Phantom | 1 | - |
46-287379G1 | - |
| Coronal/Sagittal Slice Analysis Phantom | 1 | - |
46 258559G1 | - |
| Phantom Positioner Assembly | 1 | - |
46 258709G1 | - |
|
Note:
Equipment damage possibility. Completely remove the quad head coil from the cradle before performing any body scans. Failure to do so may damage the head coil T/R network. |
| Condition | Reference | Effectivity |
|---|---|---|
|
| - | - |
About this task
Slice thickness and spatial resolution tests check the dimensional characteristics of the scanner. As this is related to the gradients being used, for the TwinSpeed the tests should be repeated for each gradient separately. In This document you will find:
HEAD SCANS
Procedure
Body Scan Procedure
Procedure
SLICE THICKNESS IMAGE ANALYSIS
Procedure
RESOLUTION IMAGE ANALYSIS
About this task
This test measures the smallest resolvable line groups in a high-contrast resolution phantom. The phantom line pair groups are oriented at a 45° angle (reference to the horizontal plane) in order to measure system resolution in the frequency and phase axes. Resolution images are acquired during the slice thickness scans, so additional scans are not required for this test.
Procedure
- Perform the head resolution check as follows:
- Repeat Step 1 to perform body resolution check. Use the Body Slice Thick exam, and display the 10-mm, L80 image (should be Series 3, Image 3).
Finalization
No finalization steps.





