o.12.6 Head Slice Thickness (3D, Grad Echo, 1.5mm)
SCAN OPTION INPUT
RX MANAGER Place cursor in Series list area.
Press right mouse button, then
Copy Series, Paste Series.
Double-click left mouse button
on new series to activate it
IMAGING PARAMETERS
Plane >Axial
Mode >3D
Pulse Seq ...GREAccept
SCAN TIMING
TR 34
Bandwidth 1.0T only 15.63
15.63 Should also be set as the default value for 1.5T and 3.0T
SCANNING RANGE
Slice Thickness 1.5
# of Scan Locs 28
ADDITIONAL PARAMETERS
Graphic Rx
See procedure for instructions.
ACQUISITION TIMING
S/I Start 0
S/I Finish 0
Freq Dir >A/P
(lowest window) Save Series
RX MANAGER Prepare to Scan
Note:
Only values different from previous series are shown. All other values should be the same as the previous series.