o.12.6 Head Slice Thickness (3D, Grad Echo, 1.5mm)

SCAN OPTION INPUT

RX MANAGER Place cursor in Series list area.

Press right mouse button, then

Copy Series, Paste Series.

Double-click left mouse button

on new series to activate it

IMAGING PARAMETERS

Plane >Axial

Mode >3D

Pulse Seq ...GREAccept

SCAN TIMING

TR 34

Bandwidth 1.0T only 15.63

15.63 Should also be set as the default value for 1.5T and 3.0T

SCANNING RANGE

Slice Thickness 1.5

# of Scan Locs 28

ADDITIONAL PARAMETERS

Graphic Rx

See procedure for instructions.

ACQUISITION TIMING

S/I Start 0

S/I Finish 0

Freq Dir >A/P

(lowest window) Save Series

RX MANAGER Prepare to Scan

Note:

Only values different from previous series are shown. All other values should be the same as the previous series.