Body Slice Thickness DataSheets

Table 1. Head Slice Thickness Data
Axial 2D Scan
Exam/Series/ImageThick(mm)LocationR1/R2/TGSystem Freq (MHz)Measured Thickness (mm)Acceptance Spec
__/ __/ __3ISOCENTER__/ __/ __ ± 0.3mm
__/ __/ __3S80 MM__/ __/ __ ± 0.3mm
__/ __/ __5ISOCENTER__/ __/ __ ± 0.5mm
__/ __/ __5S80 MM__/ __/ __ ± 0.5mm
__/ __/ __10ISOCENTER__/ __/ __ ± 1.0mm
__/ __/ __10S80 MM__/ __/ __ ± 1.0mm
Axial Echo Scan Gradient
Exam/Series/ImageThick(mm)LocationR1/R2/TGSystem Freq (MHz)Measured Thickness (mm)Acceptance Spec
__/ __/ __5ISOCENTER__/ __/ __ ± 0.5mm
Axial 3D Scan
__/ __/ __1.5ISOCENTER__/ __/ __ ± 0.1mm
Table 2. Body Slice Thickness Data
Sagital Scan
Exam/Series/ImageThick(mm)LocationR1/R2/TGSystem Freq (MHz)Measured Thickness (mm)Acceptance Spec
__/ __/ __3R80 mm__/ __/ __ ± 0.3mm
__/ __/ __3ISOCENTER__/ __/ __ ± 0.3mm
__/ __/ __5R80 mm__/ __/ __ ± 0.5mm
__/ __/ __5ISOCENTER__/ __/ __ ± 0.5mm
__/ __/ __10R80 mm__/ __/ __ ± 1.0mm
__/ __/ __10ISOCENTER__/ __/ __ ± 1.0mm
Coronal Scan
Exam/Series/ImageThick(mm)LocationR1/R2/TGSystem Freq (MHz)Measured Thickness (mm)Acceptance Spec
__/ __/ __3R80 mm__/ __/ __ ± 0.3mm
__/ __/ __3ISOCENTER__/ __/ __ ± 0.3mm
__/ __/ __5R80 mm__/ __/ __ ± 0.5mm
__/ __/ __5ISOCENTER__/ __/ __ ± 0.5mm
__/ __/ __10R80 mm__/ __/ __ ± 1.0mm
__/ __/ __10ISOCENTER__/ __/ __ ± 1.0mm
Table 3. Resolution Data
Axial Scan
Exam/Series/ImageThick(mm)LocationR1/R2/TGSystem Freq (MHz)Measured Thickness (mm)Acceptance Spec
__/ __/ __10Ri80 mm__/ __/ __ ± 0.9mm
Sagittal Scan
__/ __/ __10R80 mm__/ __/ __ ± 0.9mm