- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA307EBF20GYZ
- id_131072201.2
- Jul 5, 2019 11:19:00 PM
o.12.1 Head Slice Thickness (2D, 3mm)
SCAN OPTION INPUT
PATIENT REGISTER . New Pt
PATIENT INFORMATION
Patient Id geservice
Patient Name head slice thick
Weight (Lb)) 111
Landmark
Landmark >
Nasion
PATIENT PROTOCOLSPatient Position
PATIENT POSITION
Patient Position >Supine
Patient Entry >Head First
Coil ...Head/Neck ”GE T/R Standard Head Coil” Accept
IMAGING PARAMETERS
Plane >Axial
Mode [>] 2D
Pulse Seq [...] [Spin Echo] [Accept]
Imaging Options none
Psd Name no entry
Protocol no entry
SCAN TIMING
# of Echoes 1 (default)
TE 25
TR 300
Bandwidth 7.81 (1.0T only)
15.63 (1.5T and 3.0T)
SCANNING RANGE
FOV 20
Slice Thickness 3
Spacing 77
S/I Start I80
S/I End S80
# of Slices 3 (default)
L/R Center 0 (default)
P/A Center 0 (default)
Table Delta 0.00 (default)
ACQUISITION TIMING
Freq 256
Phase 256
NEX 4 (1.0T)
2 (1.5T and 3.0T)
Phase FOV 1.00 (default)
Freq Dir >A/P
Auto Center Freq >Peak
(lowest window) Save Series]