• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA307EBF20GYZ
  • Topic ID: id_13107220 Version: 1.2
  • Date: Jul 5, 2019 11:19:00 PM

o.12.1 Head Slice Thickness (2D, 3mm)

SCAN OPTION INPUT

PATIENT REGISTER . New Pt

PATIENT INFORMATION

Patient Id geservice

Patient Name head slice thick

Weight (Lb)) 111

Landmark

Landmark >

Nasion

PATIENT PROTOCOLSPatient Position

PATIENT POSITION

Patient Position >Supine

Patient Entry >Head First

Coil ...Head/Neck ”GE T/R Standard Head Coil” Accept

IMAGING PARAMETERS

Plane >Axial

Mode [>] 2D

Pulse Seq [...] [Spin Echo] [Accept]

Imaging Options none

Psd Name no entry

Protocol no entry

SCAN TIMING

# of Echoes 1 (default)

TE 25

TR 300

Bandwidth 7.81 (1.0T only)

15.63 (1.5T and 3.0T)

SCANNING RANGE

FOV 20

Slice Thickness 3

Spacing 77

S/I Start I80

S/I End S80

# of Slices 3 (default)

L/R Center 0 (default)

P/A Center 0 (default)

Table Delta 0.00 (default)

ACQUISITION TIMING

Freq 256

Phase 256

NEX 4 (1.0T)

2 (1.5T and 3.0T)

Phase FOV 1.00 (default)

Freq Dir >A/P

Auto Center Freq >Peak

(lowest window) Save Series]