• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30E18F20GYZ
  • Topic ID: id_13107525 Version: 3.0
  • Date: Feb 21, 2021 9:11:12 PM

Forward Reverse Quadrature Check

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable30 minutes minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
1.5T Body Loader and Phantom11.5T

2371511

-
Table 3. Safety
Note:

Equipment damage possibility. Completely remove the Quad Head Coil from the cradle before performing any body scans. Failure to do so may damage head coil T/R network.

About this task

This section determines the quadrature balance of the quad coil and the phase splitter network. In this test, the Body or Head TLT Sphere phantom is scanned with the quadrature coil connected in two ways. First, an image is taken with the quadrature coil as presently connected. Next, the cables to the quadrature switches are reversed, and the phantom is scanned a second time using the same protocol. The resultant images are analyzed to verify proper quadrature drive function and cable configuration. This procedure consist of the following:

  • Body Forward/Reverse Quadrature Scans.

  • Forward/Reverse Quadrature Test Image Analysis.

Body Forward/Reverse Quadrature Scans

Procedure

  1. At the scanner, remove any surface or head coil from the bore.
  2. Place the Body TLT Sphere Phantom in the SPT Body Loader and place the Body Loader/TLT Phantom in the center of the cradle.
  3. Landmark on the center of the sphere. On the front of the magnet enclosure, press LANDMARK on the keypad, and then press ADV TO SCAN.
  4. Start exam then select Save Rx. Then, perform scan by following steps.
    • Select Manual Prescan from the drop-down list next to the Scan button.

    • Enter the recorded center frequency value for the first installation. Verify that the center frequency peak can be seen at the center of the sceen. Then, select 'Save frequency'.

    • Select Auto Prescan from the drop-down list next to the Scan button.

    • Select the Scan .

    Figure 1. Saving Scan Parameters
  5. Reverse the black I and Q cables connecting to J1 and J2 at the bulkhead.
    Figure 2. J1 and J2 at bulkhead
  6. Without changing the prescan settings, rescan the phantom using the same scan protocol.
  7. Wait until analysis is completed before restoring the system to the original configuration (i.e., return J1 and J2 to their original connection).

Forward/Reverse Quadrative Test Image Analysis

About this task

This analysis procedure applies only to body scans.

Procedure

  1. Display each image so you can compare mean values.
  2. Use a square Region of Interest (ROI) of 4000 mm2 ± 50 mm2, located at the center of the image. Measure the mean pixel value (MPV1). In the viewer, this is shown as m=xxx. Jot down this value. See Figure 3.
    MPV1: ______________ MPV2: ______________
    Figure 3. Image Comparison
  3. Display the second image.
  4. Use a square Region of Interest (ROI) of 4000 mm2 ± 50 mm2, located at the center of the image. Measure the mean pixel value (MPV2). In the viewer, this is shown as m=xxx. Jot down this value. See Figure 3.
    Note:

    : To use the same ROI box, press CTRL-C on the keyboard for the first image. Move the mouse cursor to the second image and click the left button. Finally, press CTRL-V on the keyboard. This will copy the ROI to the second image.

  5. Use the cable configuration which produced the highest mean value from the ROI.

Finalization

Procedure

  1. Restore the system to patient scanning condition.
  2. Do a test scan to ensure the system is running properly.