o.37.6 Body Slice Thickness (Coronal, 5mm)

SCAN OPTION INPUT

RX MANAGER . Place cursor in Series list area.

Press right mouse button, then

Copy Series, Paste Series.

Double-click left mouse button

on new series to activate it.

See Note.

ADDITIONAL PARAMETERS Graphic Rx

Position cursor at P= 0 mm

This should coincide with the center of the phantom.

Start

Position cursor 80 mm away (posterior) from list location.

End

See procedure for additional information.

SCANNING RANGE

Slice Thickness 5

Spacing 75

(lowest window) Save Series

RX MANAGER Prepare to Scan

Note:

Only values different from previous series are shown. All other values Should be the same as the previous series.