o.12.2 Head Slice Thickness (2D, 5mm)
SCAN OPTION INPUT
RX MANAGER . Place cursor in Series list area.
Press right mouse button, then
Copy Series, Paste Series.
Double-click left mouse button
On new series to activate it.
See Note 1
IMAGING PARAMETERS
When using the automatic service protocol , the Imaging Options will be loaded as "none" by default. Change Imaging Options, TE, TR, and NEX as instructed below
Imaging Options ...Variable BandwidthAccept
SCAN TIMING
TE Min Full
TR 800
Bandwidth 15.63
SCANNING RANGE
Slice Thickness 5
Spacing 75
ACQUISITION TIMING
NEX 4
(lowest window) Save Series
RX MANAGER Prepare to Scan
Note:
Only values different from previous series shown. All other values should be the same as the previous series.