o.37.1 Body Slice Thickness (Sagittal, 3mm)
SCAN OPTION INPUT
PATIENT REGISTER . New Pt
PATIENT INFORMATION
Patient Id geservice
Patient Name body slice thick
Weight (Lb) 111
Landmark
Landmark >Sternal Notch
PATIENT PROTOCOLS Patient Position
PATIENT POSITION
Patient Position >Supine
Patient Entry >Head First
Coil ...BodyAccept
IMAGING PARAMETERS
Plane >Sagittal
Mode >2D
Pulse Seq ...Spin EchoAccept
Imaging Options none
Psd Name no entry
Protocol no entry
SCAN TIMING
# of Echoes 1 (default)
TE 25
TR 300
Bandwidth 7.81 (1.0T) 15.63 (1.5T and 3.0T)
SCANNING RANGE
FOV 20
Slice Thickness 3
Spacing 77
R/L Start R80
R/L End L80
# of Slices 3 (default)
P/A Center 0 (default)
I/S Center 0 (default)
Table Delta 0.00 (default)
ACQUISITION TIMING
Freq 256
Phase 256
NEX 4 (1.0T)
2 (1.5T and 3.0T)
Phase FOV 1.00 (default)
Freq Dir >A/P
Auto Center Freq >Peak
(lowest window) Save Series