• Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
  • 5690002-2EN Revision 4
  • Object ID: 00000018WIA3018DE20GYZ
  • Topic ID: id_13105865 Version: 2.0
  • Date: Feb 22, 2021 1:47:31 AM

3.0T 8-Channel Foot/Ankle Coil Setup for MCQA Test

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable15 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Small, Cylindrical Unified Phantom1Unified phantom set

5342680

-
Cubical Unified Phantom1Unified phantom set

5342681

-
Curved Adapter Panel for Flat Table (GEM)1Flat table

5395828

Sumi Motherson

Table 3. Replacement parts
ItemQuantityEffectivityPart numberManufacturer
See 3.0T HD 8-Channel Foot/Ankle Coil FRUs----
Table 4. Safety
Table 5. Required conditions
ConditionReferenceEffectivity

This coil configuration name must be installed to run the SNR test tool: HD Foot Ankle. Install this coil configuration with Config File Manager.

--

About this task

This procedure can be performed using either a flat table or a curved table.

Note:

Coils do not ship with phantoms. Phantoms come in a unified phantom set with the MR system.

Topic ID: id_SL5071411-1104206

Unified Phantom Setup

About this task

Follow this process to prepare for the automated SNR test using the 3.0T HD 8-channel foot/ankle coil by Invivo (M3340CB).

Procedure

  1. (For flat table) Insert the curved adapter panel into the magnet end of the table.
    Figure 1. Curved Adapter Panel Assembly on Flat Table
  2. Place the foot/ankle coil baseplate on the table.
    Figure 2. Foot/Ankle Coil Baseplate on Curved Table
    Figure 3. Foot/Ankle Coil Baseplate on Flat Table
  3. (For curved table) Place the small cylindrical unified phantom horizontally on the baseplate.
    Figure 4. Cylindrical Unified Phantom on Baseplate on Curved Table
  4. Place the cubical unified phantom vertically. Secure the phantom to the baseplate using the Velcro strap.
    Figure 5. Cubical Unified Phantom on Baseplate
  5. Cover the baseplate with the coil.
    Figure 6. Coil over Baseplate
  6. Connect the foot/ankle coil to port A on the LPCA. Landmark the coil at the crossmark on the coil handle.
    Figure 7. Crossmark on Coil Handle
  7. Advance to scan.
  8. Perform the Multi-Coil Quality Assurance Tool procedure.
  9. After the test is complete, carefully remove the cubical phantom from the baseplate. Then carefully remove the cylindrical phantom. Avoid damage to the phantoms.
Topic ID: id_14930126

Finalization

Finalization

No finalization steps.