Checking the GOC battery replacement

Check the battery replacement sticker on the GOC.

Prerequisites

Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 5 minutes -
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Check the battery replacement sticker on the GOC.
    The sticker should be attached to lower left or right side of the GOC. Batteries should be replaced every three years.
    • If the date on the battery replacement sticker is within 12 months, order battery and replace during the next PM per the Lower Level FRU Battery Replacement procedure for that computer model.
    • If the GOC does not have a sticker, order a battery and replace during the next PM per the Lower Level FRU Battery Replacement procedure for that computer model.