• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA3077D010GYZ
  • Topic ID: id_2002986 Version: 3.4
  • Date: Jul 7, 2021 3:59:54 PM

Removing the magnet right-hand center cover

Removes the center cover from the right-hand side of the magnet.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2-10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
The right hand top cover is removed.
The right hand bottom cover is removed.

Procedure

  1. Remove three screws securing the right-hand center cover to the magnet.
    Figure 1. Right-hand center cover screws
  2. Lift the cover up and separate the cover slightly from the magnet.
  3. Disconnect the M3350 cable connector from the enclosure ambient temperature sensor.
  4. Disconnect the M3379 cable connector from the enclosure LED switch J60.
  5. Remove the cover from the magnet.