• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30379350GYZ
  • Topic ID: id_2020661 Version: 1.6
  • Date: Jul 7, 2021 3:51:54 PM

Removing the dockable table interface cover

Removes the dockable table interface cover from the magnet front cover.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
Left- and right-hand lower access panels are removed.

Procedure

  1. If docked, raise the patient table to the up-limit position.
  2. Remove the four screws securing the dockable table interface cover to the magnet front cover.
    Figure 1. Dockable table interface cover screws
  3. Remove the dockable table interface cover from the magnet front cover.
    Figure 2. Dockable table interface cover