Replacing the DS Digitized Per Pin (DPP) Receiver Distributor (Rx Dist)
Removal and installation instructions for replacing the DS Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) in the magnet outer electronics.
Prerequisites
Gen 2 DS DPP RRx and Gen 2 DS RxDist requires PX28 Software.
The following tables represent a comprehensive roll-up of the personnel, time, tools, test equipment, and replacement parts you need to complete all possible tasks in this procedure.
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Digital Voltmeter (DVM) | 1 | 46-194427P284 | - |
| Red Warning LOTO Tag | 2 | 46-194427P322 | - |
| Multi-Locking Device (if multiple service personnel are involved) | 1 | 46-194427P313 | - |
| Red LOTO Personal Lock Wrap | 2 | 2393068 | - |
| Brass Master Padlock with Brass Shackle | 2 | 46-194427P320 | - |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 Kit | 5113258 | - |
| Nonmagnetic Titanium Service Tool Kit, Large Set | 5112581 | - | |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| DS Rx Distribution HLA Note: DS DPP requires PX28 Software. | 1 | 5434906-2 | - |
| Required conditions |
|---|
| System Power must be turned OFF. See Lockout/Tagout for ISC. |
| Safety | ||||||||
|---|---|---|---|---|---|---|---|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. | ||||||||
| ||||||||