• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA303772GYZ
  • Topic ID: id_2001617 Version: 2.1
  • Date: Aug 17, 2021 6:31:49 PM

Replacing the bore light

Removes a bore light from the body coil and installs a replacement bore light.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2-1 hour and 35 minutes -
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Consumables
ItemQuantityPart numberManufacturer
3.6 inch Cable Tie246-208758P1-
Replacement parts
ItemQuantityPart numberManufacturer
Bore Light Assembly FRU15561809-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

There are four bore light assemblies, one on each side of the body coil at each end of the magnet bore. This procedure is applicable to any of the four assemblies.