Finalizing the 1-piece bridge replacement
This procedure finalizes the 1-piece bridge replacement.
Procedure
- Removing LOTO - ISC.
- Doing a check scan.
- Starting up the host computer.
- Make sure the scan setting is correct. See Setting the patient table scan range configuration.
- Make sure the cradle moves in and out smoothly.
- Do a cradle motion test. See Doing the cradle motion test.