Finalizing the 1-piece bridge replacement

This procedure finalizes the 1-piece bridge replacement.

Procedure

  1. Removing LOTO - ISC.
  2. Doing a check scan.
  3. Starting up the host computer.
  4. Make sure the scan setting is correct. See Setting the patient table scan range configuration.
  5. Make sure the cradle moves in and out smoothly.
  6. Do a cradle motion test. See Doing the cradle motion test.