• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA309B62GYZ
  • Topic ID: id_2001431 Version: 5.0
  • Date: Apr 25, 2019 2:11:53 AM

Checking the body coil air flow

Checks the function of the body coil air blower system.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Required conditions
System software must be booted.

Procedure

  1. Open Service Browser and select Diagnostics > System Function > Patient Handling > SRI Functional Diagnostics > .
  2. Select SRI Real-Time Data and click Run.

    Result

    Air Flow Sensor 0 = ##### 0.1 FPM should be displayed in the Test Results field.
  3. Confirm that the air flow feedback is greater than 440 fpm (= 4400 0.1 FPM in Test Results).
    Feedback displayedResult
    Less than 440 fpmThe system has failed the check.
    Greater than or equal to the normal mode set pointThe system has passed the check and is functioning properly.