Discovery and Optima Front Status Indicators
Overview
This document describes the state of external status indicators, such as LEDs or lamps, visible from exterior or front of a component while system is in idle or standby state. With exception of LPCA, this document does not describe indicators inside component case or indicators on component circuit board. Indicators listed are those visible without need to withdraw component from chassis or to remove component covers (except for LPCA).
The document is divided into sections by cabinet or high-level component. Each section contains a table that explains the status indicators on each component. The tables are divided into 3 columns.
-
The first column on left is labeled “Front Status Indicators” and lists the printed name on the front panel for the indicator, if available, or identifies it by description if it is unlabeled.
-
The second column in middle is labeled “Status” and describes the indicator status when the system is idle or in standby condition. Indicators that can display multiple normal states may have one color or another listed in this column.
-
The third column on right is labeled “Indicator Description”. This column describes the indicator states and condition indicated by each as well as location detail and other helpful information.
Be cautious of dim LEDs. Light from nearby lit, active LEDs often “bleeds” into inactive, dark LEDs making inactive LEDs appear active.
Power Gradient Rf cabinet - Section 2
-
XGA – Table 1
-
XPS – Table 2
-
XG2 – Table 3
-
XPS2 – Table 4
-
Cronus Chassis GMC Board – Table 5
-
Cronus Chassis XGA Control Board – Table 6
-
Cronus Chassis XPS Control Board – Table 7
-
Cronus Chassis Power Supply Board – Table 8
-
4100 ICN – Table 9
-
4170 ICN – Table 10
-
Gen 5 & 6 ICN – Table 11
-
AUX Module – Table 12
-
PEM Module – Table 13
-
CAM AGP2 Board – Table 14
-
CAM IRF3 Board – Table 15
-
CAM IXG Board – Table 16
-
CAM SCP3 Board – Table 17
-
CAM SRF/TRF Board – Table 18
-
CAM PSE board – Table 19
-
CAM PSE I/O board – Table 20
-
CAM ERF board – Table 21
-
CAM Power Supply Board – Table 22
-
CAM NB RF AIF Board – Table 23
-
CAM BB RF AIF Board – Table 24
-
CAM RF Detector Board – Table 25
-
CAM RF Processor Board – Table 26
-
CAM Chassis – Table 27
-
ICE Chassis – Table 28
-
Ethernet Switch – Table 29
-
Term Server – Table 30
-
1.5T SRFD2 – Table 31
-
1.5T XRFD – Table 32
-
3T XRFD single-drive – Table 33
-
3T XRFD dual-drive – Table 34
-
PDU – Table 35
Penetration Cabinet - Section 3
-
CFB – Table 36
-
DTX1 NB Exciter – Table 37
-
DTX2 BB Exciter (optional) – Table 38
-
LED PS – Table 39 (no indicators)
-
HOS (optional) – Table 40
-
PHPS2 – Table 41
-
Driver Module – Table 42
-
SRPS – Table 43
-
eSRPS – Table 44
-
AC Distribution Module – Table 45
Heat Exchanger Cabinet – Section 4
Heat Exchanger Cabinet – Table 46
Rear Pedestal - Section 5
-
RF Hub MCDB – Table 50
-
RF Hub RFCB – Table 51
-
RF Hub ePDB – Table 54
-
Table Interface Board – Table 55 (no indicators)
-
Reroute Box – Table 56 (no indicators)
-
Long Drive Motor Assembly – Table 57 (no indicators)
-
LPCA – Table 58
Magnet - Section 6
-
TDM – Table 59 (no indicators)
-
Body Hybrid Splitter – Table 60 (no indicators)
-
PAC – Table 61 (no indicators)
-
RRx – Table 62
-
DPP RRx – Table 63
-
DPP RxDist – Table 64
-
SRI – Table 65
-
DTRSW – Table 66 (no indicators)
-
Resistor Splitter Box – Table 67 (no indicators)
-
IRD Monitor – Table 68 (no indicators)
-
Enclosure Operator Console Panels – Table 69 (no indicators)
-
Dock – Table 70 (no indicators)
Patient Table – Section 7 (no indicators)
Patient Table – Table 71
Magnet Monitor 3 - Section 8
Magnet Monitor 3 – Table 72
Global Operator’s Console - Section 9
-
Host PC – Table 73
-
Ethernet Switch – Table 74
-
GOC USB-to-Serial Converter Module – Table 75
-
GOCAA – Table 76 (no indicators)
-
IRD Converter Module – Table 77
F-50 Shield Cooler Compressor – Section 10
F-50 Shield Cooler Compressor – Table 78
Operator’s Workspace - Section 11
-
Host Monitor – Table 79
-
SCIM – Table 80
-
Keyboard – Table 81
-
Mouse – Table 82 (no indicators)
-
O2 Monitor ISA-40M – Table 83 (optional)
-
O2 Monitor ISA-60M & MRI-5175 – Table 84 (optional)
-
O2 Monitor ISA-42M – Table 85 (optional)
MNS Option - Section 12
MRU – Section 13
Magnet Rundown Unit, all GE Magnet types – Table 88
MDP – Section 14
Teal, GEXPRO, or BEVCO Main Disconnect Panel – Table 89
PET/MR - Section 15
-
Magnet Room Leak Sensor Probe – Table 90
-
Magnet Room CPDA – Table 91
-
PET Cabinet DMOD – Table 92
-
PET Cabinet 80V PS – Table 93
-
PET Cabinet Fiber Optic Breakout Box – Table 94 (no indicators)
-
PET Cabinet PROP – Table 95
-
PET Cabinet 16-Port Ethernet Switch – Table 96
-
PET Cabinet Computer – Table 97
-
PET Chiller – Table 98
-
PET Cabinet PDU – Table 99
Glossary – Section 16
Glossary – Table 100
PGR Cabinet
| Front Status Indicators | Status | Indicator Description |
| OCT POWER | Green | Indicates power on Octavius link. |
| OCT STATUS | Yellow or green | Indicates activity on Octavius link. |
| LL0L | Red to green | Lv_Left_0_Lower. Red – IGBT gated on; Green – IGBT standby, off. |
| LV0 OUT | Dark or red to green | Low_Voltage_0_OUT. Red – bridge voltage present; Green – reverse bridge voltage present; Dark – no voltage present. |
| LR0U | Red to green | Lv_Right_0_Upper. Red – IGBT gated on; Green – IGBT standby, off. |
| XGA FAULT | Dark | Orange indicates fault in XGA. |
| COMM FAULT | Dark | Orange indicates communication fault at J2 SFP interface. Does not indicate hardware failure. |
| ENABLED | Dark | Green indicates XGA power electronics enabled. |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| Front Status Indicators | Status | Indicator Description |
|---|---|---|
| OCT POWER | Green | Indicates power on Octavius link. |
| OCT STATUS | Yellow or green | Indicates activity on Octavius link. |
| COMM FAULT | Dark | Yellow indicates communication fault at J5 SFP interface. Does not indicate hardware failure. |
| ENABLED | Dark | Green indicates XGA power electronics enabled. |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| XPS FAULT | Dark | Orange indicates fault in XPS. |
| Front Status Indicators | Status | Indicator Description |
|---|---|---|
| LL0L | Red to green | Lv_Left_0_Lower. Red – IGBT gated on; Green – IGBT standby, off. |
| LV0 OUT | Red to green | Low_Voltage_0_OUT. Red – bridge voltage present; Green – reverse bridge voltage present; Dark – no voltage present. |
| LR0U | Red to green | Lv_Right_0_Upper. Red – IGBT gated on; Green – IGBT standby, off. |
| X-AXIS | Dark or Green | Green indicates XGA performing X-axis functions as set by Octavius. |
| Z-AXIS | Dark or Green | Green indicates XGA performing Z-axis functions as set by Octavius. |
| Y-AXIS | Dark or Green | Green indicates XGA performing Y-axis functions as set by Octavius. |
| LINK-UP | Green | Indicates Octavius link active and cable connected at J12. |
| ERROR | Dark | Red indicates XGA fault condition. |
| WARNING | Dark or Yellow | Yellow indicates that a parameter has trended out of limit, but alone this will not stop scanning and must see message log for explanation. |
| ENABLED | Dark | Green indicates XGA power electronics enabled. |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| Front Status Indicators | Status | Indicator Description |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| ENABLED | Dark | Green indicates XPS power electronics enabled. |
| WARNING | Dark or Yellow | Yellow indicates that a parameter has trended out of limit, but alone this will not stop scanning and must see message log for explanation. |
| ERROR | Dark | Red indicates XPS fault condition. |
| Y AXIS | Dark or Green | Green indicates XGA performing Y-axis functions as set by Octavius link. |
| LINKUP | Green | Indicates Octavius link active and cable connected at J1. |
| Z AXIS | Dark or Green | Green indicates XGA performing Z-axis functions as set by Octavius link. |
| X AXIS | Dark or Green | Green indicates XGA performing X-axis functions as set by Octavius link. |
| Front Status Indicators | Status | Indicator Description |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| READY | Dark | Green indicates XGD power electronics enabled. |
| LINK FLT | Dark or Amber | Amber indicates any fiber optic communication failure. |
| GRAD FLT | Dark | Amber indicates fault in XGD subsystem. |
| Ethernet link/activity | Yellow | On is link valid. Flashing indicates activity at any speed. |
| Ethernet link 1000/activity | Green | On is 1000 link valid. Flashing indicates 1000 activity. |
| Front Status Indicators | Status | Indicator Description |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| LINK FLT | Dark or Amber | Amber indicates failure on fiber optic to any XGA. |
| XGA FLT | Dark or Amber | Amber indicates fault reported from XGA. |
| READY | Dark | Green indicates XGA ready and enabled. |
| Front Status Indicators | Status | Indicator Description |
| HEARTBEAT | Green | Flashing indicates processor executing. Successful TPS Reset required. |
| LINK FLT | Dark or Amber | Amber indicates failure on fiber optic to any XPS. |
| XPS FLT | Dark or Amber | Amber indicates fault reported from XPS. |
| READY | Dark | Green indicates XPS ready and enabled. |
| Front Status Indicators | Status | Indicator Description |
| PWR GOOD | Green | Power supply operation normal. |
| FAULT | Dark | Yellow indicates power supply fault, often internal. |
| Front Status Indicators | Location | Status | Indicator Description |
| REAR PS | ICN chassis | Dark | Yellow when power supply monitoring limits out of tolerance. May not initially coincide with ICN malfunction. |
| Temperature symbol | Dark | Yellow if ICN goes over temperature. | |
| TOP FAN | Dark | Red to indicate incorrect internal fan rotation speed. | |
| Arrow at top of concentric circles | Dark | Locate LED/Button to light green and identify ICN; push to light and manually identify ICN or can be activated remotely to light and identify ICN. | |
| Exclamation symbol inside triangle | Dark | Red indicates an internal ICN fault (bad fan, memory map issue) requiring service. | |
| OK | Green | Lit when unit functioning. | |
| Block icon | Hard drive | Dark | Blue indicates hard drive ready for removal as part of service activity. |
| Exclamation symbol inside triangle | Dark | Red when hard drive reporting fault. | |
| OK | Green | Lit when hard drive functioning. |
| Front Status Indicators | Location | Status | Indicator Description |
| REAR PS | ICN chassis | Dark | Red when power supply monitoring limits out of tolerance. May not initially coincide with ICN malfunction. |
| Temperature symbol | Dark | Red if ICN goes over temperature. | |
| TOP FAN | Dark | Red to indicate incorrect internal fan rotation speed. | |
| Arrow at top of concentric circles | Dark | Locate LED/Button to light green and identify ICN; push to light and manually identify ICN or can be activated remotely to light and identify ICN. | |
| Exclamation symbol inside triangle | Dark | Yellow indicates an internal ICN fault (bad fan, memory map issue) requiring service. | |
| OK | Green | Lit when ICN functioning. | |
| Block icon | Hard drive | Dark | Blue indicates hard drive ready for removal as part of service activity. |
| Exclamation symbol inside triangle | Dark | Red when hard drive reporting fault. | |
| OK | Green | Lit when hard drive functioning. |
| Front Status Indicators | Location | Status | Indicator Description |
| Circle with vertical line power button | ICN chassis | Dark | Lit when ICN receiving power and energized. |
| Unlabeled button to left of USB port | Dark or Green | Locate LED/Button to light green and identify ICN; push to light and manually identify ICN or can be activated remotely to light and identify ICN. | |
| Sine squiggle symbol | Hard drive | Green | Lit when hard drive receiving power and active. |
| Cylinder symbol | Dark or Green | Flashes green to indicate hard drive activity. |
| Front Status Indicators | Status | Indicator Description |
| FAULT | Dark | Yellow indicates fault interrupting Aux module operation. May be external and not internal. |
| HEARTBEAT | Green | Flashing indicates processor operational. |
| POWER | Green | Indicates internal 3.3V power regulator OK. |
| Oct power | Green | Indicates Octavius cable connected. |
| Oct status | Green | Indicates Octavius link activity. |
| Front Status Indicators | Status | Indicator Description |
| NIOS HB | Green | After PEM download from successful TPS Reset flashing LED indicates NIOS processor OK. |
| POWER OK | Green | PEM incoming power, 3.3V regulator OK. |
| FAULT | Dark | Red indicates fault condition; usually external but can be internal. |
| E STOP DET | Dark | Green indicates E-Stop condition detected from pressed E-Stop button or other condition that causes E-Stop. |
| E STOP CONTROL FPGA | Dark | Green indicates PEM caused E-Stop action; gradient or PDU fan failure or failure of PEM code to install during TPS Reset. |
| E OFF 12V | Green | Normally lit when system operational and E-Off is inactive. |
| Ethernet link/activity | Yellow | On is link valid. Flashing indicates activity at any speed. |
| Ethernet link 1000/activity | Green | On is 1000 link valid. Flashing indicates 1000 activity. |
| Front Status Indicators | Status | Indicator Description |
| CPU | Green | Flashes in proportion to CPU activity. |
| FAIL | Dark | Red indicates fault condition, usually due to external factor outside unit. Do not replace board based on LED status alone. |
| Front Status Indicators | Status | Indicator Description |
| RST REQ | Dark or green | Green indicates system manual reset request from pressing front push button. |
| PWR OK | Green | Power good status from onboard regulator OK. |
| J14 RDY | Dark or green | Green indicates socket in use, DVMR link up. |
| J13 RDY | ||
| J12 RDY | ||
| J11 RDY | ||
| J10 RDY | ||
| REF OK | Green | Steady green indicates clock signal from DTX1 at J8 is OK. Flashing indicates clock signal missing. |
| Front Status Indicators | Status | Indicator Description |
| POWER | Green | Power good for IXG. |
| REF | Green | Indicates clock signal from DTX1 at J8 is OK. Flashing indicates clock signal missing. |
| READY | Dark | Green LED but normally dark; not used. |
| HB | Green | XGP NIOS Heartbeat; may not flash until after successful download. |
| J11 READY | Dark or Green | Green when socket in use and DVMR Link up. |
| J11 FAULT | Dark or Red | Red when fault on DVMR Link. |
| J12 READY | Dark or Green | Green when socket in use and DVMR Link up. |
| J12 FAULT | Dark or Red | Red when fault on DVMR Link. |
| J13 READY | Dark or Green | Green when socket in use and DVMR Link up. |
| J13 FAULT | Dark or Red | Red when fault on DVMR Link. |
| J14 READY | Dark or Green | Green when socket in use and DVMR Link up. |
| J14 FAULT | Dark or Red | Red when fault on DVMR Link. |
| J15 READY | Dark or Green | Green when socket in use, XG2 X link up. |
| J15 FAULT | Dark or Red | Red when fault on XG2 X interface. |
| J16 READY | Dark or Green | Green when socket in use, XG2 Y link up. |
| J16 FAULT | Dark or Red | Red when fault on XG2 Y interface. |
| J17 READY | Dark or Green | Green when socket in use, XG2 Z link up. |
| J17 FAULT | Dark or Red | Red when fault on XG2 Z interface. |
| J18 READY | Dark or Green | Green when socket in use, XPS2 X link up. |
| J18 FAULT | Dark or Red | Red when fault on XPS2 X link. |
| J19 READY | Dark or Green | Green when socket in use, XPS2 Y link up. |
| J19 FAULT | Dark or Red | Red when fault on XPS2 Y link. |
| J20 READY | Dark or Green | Green when socket in use, XPS2 Z link up. |
| J20 FAULT | Dark or Red | Red when fault on XPS2 Z link. |
| Ethernet link 1000/activity | Green | On is 1000 link valid. Flashing indicates 1000 activity. |
| Ethernet link/activity | Yellow | On is link valid. Flashing indicates activity at any speed. |
| Front Status Indicators | Status | Indicator Description |
| S1 | Green | Lit to indicate SCP3 OS initialization started. |
| S2 | Dark | Yellow LED used to indicate an operational condition but normally dark. Not used. |
| S3 | Dark | |
| FAULT | Dark or Red | Red indicates a software controlled ‘fault’. Not an indication of SCP3 board condition. Do not replace SCP3 based on state of this LED. |
| HS | Dark | Blue LED service indicator but normally dark. Not used. |
| PWR | Green | Indicates onboard voltage regulators OK. |
| Ethernet link/activity | Yellow | On is link valid. Flashing indicates activity at any speed. |
| Ethernet link 1000/activity | Green | On is 1000 link valid. Flashing indicates 1000 activity. |
| Front Status Indicators | Status | Indicator Description |
| 1.8V | Green | SRF; onboard 1.8V onboard power regulator OK. |
| 2.5V | SRF; onboard 2.5V onboard power regulator OK. | |
| DSP0 | SRF; flashing DSP0 (SSP, X, Y, Z sequencers) after successful TPS Reset. | |
| DSP1 | SRF; flashing DSP1 (Rho1, Rho2, Theta1, Theta2 sequencers) after successful TPS Reset. | |
| 1.9V | TRF; onboard 1.9V onboard power regulator OK. | |
| WARP | TRF; flashing WARP after successful TPS Reset. | |
| SPU | TRF; SPU DUART Ch. A - waveform display data active; often inactive and LED dark. |
| Front Status Indicators | Status | Indicator Description |
| HOST RST | Dark or green | Green indicates processor has detected a host reset event caused by either a TPS Reset or the Chassis Reset button pressed. |
| PWR OK | All onboard power supplies OK. | |
| SPU HB | Green | Flashing after successful TPS Reset indicating SPU processor is executing normally. |
| WARP HB | Flashing after successful TPS Reset indicating WARP processor is executing normally. | |
| NIOS HB | Flashing after successful TPS Reset indicating NIOS processor is executing normally. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| PWR OK | Green | Lit when incoming power OK. |
| J14 RDY | Dark or Green | Green when socket in use, DVMR Link up. |
| J13 RDY | Green when socket in use, DVMR Link up. | |
| J12 RDY | Green when socket in use, DVMR Link up. | |
| J11 RDY | Green when socket in use, DVMR Link up. | |
| J10 RDY | Green when socket in use, DVMR Link up. | |
| REF OK | Green | Clock signal originating from DTX1 at J8 OK. Flashing indicates clock signal missing. |
| Front Status Indicators | Status | Indicator Description |
| POWER GOOD | Green | Power good status. |
| FAULT | Dark | Yellow indicates internal board fault. |
| Front Status Indicators | Status | Indicator Description |
| 1.5V | Green | Internal 1.5V power present. |
| CAN PWR | Green | +24V CAN bias present. |
| CAN STAT | Green | Flashes in 2 ON, 2 OFF pattern to indicate normal operation. |
| RF DISABLE | Dark or Yellow | Yellow indicates fault condition, unblank inhibited; condition usually external to unit. |
| Front Status Indicators | Status | Indicator Description |
| 1.5V | Green | Internal 1.5V power present. |
| CAN PWR | Green | +24V CAN bias present. |
| CAN STAT | Green | Flashes in 2 ON, 2 OFF pattern to indicate normal operation. |
| RF DISABLE | Dark or Yellow | Yellow indicates fault condition, unblank inhibited; condition usually external to unit. |
| Front Status Indicators | Status | Indicator Description |
| SENSE | Dark or Yellow | One LED for each channel. Yellow when body, head, or MNS RF transmit energy, minimum 10 Watts, detected. |
| Front Status Indicators | Status | Indicator Description |
| 1.5V | Green | Internal 1.5V power present. |
| CAN PWR | Green | +24V CAN bias present. |
| CAN STAT | Green | Flashes in 2 ON, 2 OFF pattern to indicate normal operation. |
| INHIBIT | Dark or Yellow | Yellow indicates UPM fault condition inhibiting unblank; condition usually external to unit and often not a hardware failure. |
| Front Status Indicators | Status | Indicator Description |
| BLOWER GOOD (left) | Green | Lit if fan tach normal, fan rotating at correct speed. |
| BLOWER GOOD (right) | Green | Lit if fan tach normal, fan rotating at correct speed. |
| Front Status Indicators | Status | Indicator Description |
| Up or down-pointing arrows between SFPs | Green or dark | Arrow direction points to indicated SFP. Green LEDs indicate link up. Dark LED indicates link inactive or SFP normal. Lit red indicates SFP status abnormal. |
| USB | Green | Pioneer: Host PC power ON. Non-Pioneer: USB Hub power ON. |
| CAN | Green | Flashes with CAN activity. Dark is CAN disconnected, no activity. |
| PWR | Green | Indicates internal mainboard power good. |
| INPUT PWR | Green | Indicates 48VDC PS OK. |
| CLK | Green |
Indicates the status of the ICE reference clock signal at J28.
|
| HB | Green |
Indicates heartbeat from ASP (AGP SCP Processor) on ICE.
|
| ICEFL | Dark | Red indicates failure in ICE internal POST event process OR ICE in reset. |
| FANFL | Dark | Red indicates fault of fan on rear of ICE OR ICE in reset. |
| IDB PWR | Green | Indicates Integrated Daughter Board mainboard power OK. Not present on later models. |
| Ethernet link/activity | Yellow | J53 & J54; On is link valid. Flashing indicates activity at any speed. |
| Ethernet link 1000/activity | Green | J53 & J54; On is 1000 link valid. Flashing indicates 1000 activity. |
| Front Status Indicators | Status | Indicator Description |
| Power | Green | Lit to indicate normal power supply operation. |
| 1-8 or 1-16 or 1-24 or 1-32 | Green or yellow or dark | Solid green indicates 1Gbit link valid. Flashing green indicates activity at 1Gbit speed. Solid yellow indicates link valid at 10/100 speed. Flashing yellow indicates activity at 10/100 speed. Dark indicates port unused or inactive. |
| Front Status Indicators | Status | Indicator Description |
| LINK | Green | On is 1000 link valid. Flashing indicates 1000 activity. Located in upper left furthest from reset button. |
| ACT | Yellow | On is link valid. Flashing indicates activity at any speed. Located in lower left furthest from reset button. |
| PWR | Green | Indicates unit receiving power. Located in upper right closest to reset button |
| READY | Dark | During configuration flashes yellow to indicate unit status. Located in lower right closest to reset button. |
| Front Status Indicators | Status | Indicator Description |
| DC HV | Red | HV supply active/not discharged to a safe level; normal operating state. Dark indicates HV supply inactive/discharged to a safe level. |
| UNBLANK | Dark | Flashes yellow to indicate when unblank signal active. |
| TEST | Dark | Green indicates amp in test mode. Sometimes used during HSS. |
| HEAD | Dark or Green | Green indicates amp in head mode. |
| BODY | Dark or Green | Green indicates amp in body mode. |
| PMON | Dark | Red indicates power monitor fault. |
| AC POWER | Green | AC power to amp is present. |
| POWER OFF | Dark | Yellow indicates amp in power off state although incoming AC power present. |
| WAIT | Dark | Yellow indicates amp in transition state (processing command). |
| STANDBY | Dark | Green indicates amp in standby state. |
| OPERATE | Dark | Green indicates amp in operate state. |
| FAULT | Dark | Red indicates failure event causing amp to stop pulsing. Failure is often external to amp. |
| OFF | Dark | Green when unit in power off state. |
| STBY | Dark | Green when unit in standby state. |
| WAIT | Dark | Yellow when unit in transient state such as processing a command. |
| OPER | Dark | Green when unit in operate state. |
| RFLK | Dark | Yellow when system asserts RFLOCK signal. |
| FLT | Dark | Red indicates internal amplifier fault condition. |
| PWR | Yellow | AC line power to unit present. |
| DCHV | Green | Green when HV supply for XRFD active/not discharged to safe level. |
| UNBL | Dark | Yellow when unblank signal active and present. |
| TEST | Dark | Green when unit in TEST mode. Sometimes used during HSS. |
| BODY | Dark or green | Green when unit in BODY mode. |
| HEAD | Dark or green | Green when unit in HEAD mode. |
| Front Status Indicators | Status | Indicator Description |
| FAULT | Dark | Red for internal amplifier fault condition. |
| RF LOCK | Dark | Yellow when system asserts RFLOCK signal. |
| OPERATE | Dark | Green when unit in operate state. |
| WAIT | Dark | Yellow when unit in transient state such as processing a command. |
| STANDBY | Dark | Green when unit in standby state. |
| POWER OFF | Dark | Yellow when unit in power off state. |
| HEAD | Dark or green | Green when unit in HEAD mode. |
| BODY | Dark or green | Green when unit in BODY mode. |
| TEST | Dark | Green when unit in TEST mode. Sometimes used during HSS. |
| UNBLANK | Dark | Yellow when unblank signal active and present. |
| DC HIGH VOLTAGE 1 | Green | Green when HV supply 1 for XRFD active/not discharged to safe level. |
| DC HIGH VOLTAGE 2 | Green | Green when HV supply 2 for XRFD active/not discharged to safe level. |
| AC POWER | Green | AC line power to unit present. |
| Front Status Indicators | Status | Indicator Description |
| FAULT | Dark | Orange when internal fault of unit on any channel. |
| RFLOCK | Dark | Green when system asserted RF lock signal on both channels. |
| OPERATE | Dark | Green when unit in operate state for both channels. |
| UNBLANK | Dark | Yellow when unblank signal present and active for both channels. |
| STANDBY | Dark | Green when unit in standby state for both channels. |
| POWER OFF | Yellow | Yellow when unit in power off state for both channels. |
| HEAD | Dark or green | Green when unit in HEAD mode, channel 1. |
| BODY | Dark or green | Green when unit in BODY mode, both channels. |
| TEST | Dark | Green when unit in TEST mode, channel 1. Sometimes used during HSS. |
| WAIT | Dark | Yellow when unit in transient state such as processing a command, both channels. |
| DC HIGH VOLTAGE | Green | Green when HV supply for XRFD active/not discharged to safe level. |
| AC POWER | Green | AC line power to unit present. |
| Front Status Indicators | Status | Indicator Description |
| ɸA | Green | Lit when incoming AC power on phase A and main 80 kVA PDU circuit breaker closed/engaged. |
| ɸB | Green | Lit when incoming AC power on phase B and main 80 kVA PDU circuit breaker closed/engaged. |
| ɸC | Green | Lit when incoming AC power on phase C and main 80 kVA PDU circuit breaker closed/engaged. |
Penetration Cabinet
| Front Status Indicators | Status | Indicator Description |
| CAN STAT | Green | Flashes in 2 ON, 2 OFF pattern when CAN bus normal. |
| CAN PWR | Green | Lit when CAN +24V bias present. |
| PWR OK | Green | Lit when all digital power supplies OK. |
| HRT BEAT | Green | Flashing when NIOS processor executing. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| HEARTBEAT | Green | Flashes indicating CPU operating. |
| Comm Err: | Dark | Red when CAN +24V bias missing. |
| Comm Stat: | Green | Flashes in 2 ON, 2 OFF pattern to reflect normal CAN bus operation. |
| Power OK: | Green | Lit when internal power normal. |
| Func Err: | Dark | Red when unit reports internal fault. |
| Out Enable: | Dark or green | Green when current output through coil windings. |
| Watchdog: | Green | Normally lit to indicate watchdog enabled. |
| Heartbeat: | Green | Flashes indicating FPGA active. |
| 5V Stat: | Green | Lit when 5V power enabled. |
| Front Status Indicators | Status | Indicator Description |
| PHPS2 POWER | Green | Front switch backlit when in ‘I’ position and AC power present. |
| Front Status Indicators | Status | Indicator Description |
| HV DISABLE | Dark | Yellow when SW1 in down disabled position. |
| TR DISABLE | Dark | Yellow when SW2 in down disabled position. |
| DD DISABLE | Dark | Yellow when SW3 in down disabled position. |
| UNUSED | Dark | Yellow when switch marked ‘Unused’ in down disabled position. |
| CCC ERROR | Dark | Red when CAN Core Comm board detects failure on CAN link. |
| CCC HRT | Green | Flashing to indicate normal CAN Core Comm heartbeat. |
| MODULE ERROR | Dark | Red to indicate failure internal to Driver Module. |
| CCC STATUS | Green | Flashes in 2 ON, 2 OFF pattern when CAN link operation normal. |
| DCB HRT | Green | Flashing to indicate normal Driver Cntrl Board heartbeat. |
| DRIVER FAULT | Dark | Red when module operation interrupted. Can correlate with INIHIBIT LED status. Most of the time does NOT imply a hardware fault. |
| NB UNBLANK | Dark or green | Flashing green when NB unblank active. |
| BB UNBLANK | Dark or green | Flashing green when BB unblank active. |
| CW UNBLANK | Dark or green | Flashing green when CW unblank active. |
| INHIBIT | Dark or red | Red when inhibit line active. Alone does NOT indicate hardware failure. |
| Front Status Indicators | Status | Indicator Description |
| PHASE A | Green | Only present on early units. Indicates incoming AC for phase A. |
| PHASE B | Green | Only present on early units. Indicates incoming AC for phase B. |
| PHASE C | Green | Only present on early units. Indicates incoming AC for phase C. |
| FAULT | Dark | Located on SRPS Control Board. Red when fans faulted or internal power supplies disabled due to open cable, other fault condition. |
| Front Status Indicators | Status | Indicator Description |
| eSRPS POWER | Green | Lit when incoming AC to unit OK. |
| EXCITER POWER | Green | Lit when EXCITER receiving power. Blinking indicates fault. |
| MNS EXCITER POWER | Dark or green | Green when MNS receiving power. Dark if MNS option not installed. Blinking indicates fault. |
| RF HUB POWER | Green | Lit when RF Hub or Rx Distribution module receiving power. Blinking indicates fault. |
| RF HUB CABLE | Green | Lit when RF Hub or Rx Distribution module cable connected at J1, confidence loop OK. Blinking indicates fault. Dark indicates inactive. |
| TDM CABLE | Green | Lit when TDM cable or Rx Distribution module connected, confidence loop OK. Blinking indicates fault. |
| TDM POWER GOOD | Green | Lit when TDM or Rx Distribution module receiving power. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
Heat Exchanger Cabinet
| Front Status Indicators | Status | Indicator Description |
| VFD GPMP | Red or LCD | Displayed value varies depending standby or ready state and pump type. See HEC Electrical Box replacement document for details. Variation of a few units OK. |
| VFD PPMP | Red or LCD | Displayed value varies depending in standby or ready state and hardware in PGR cabinet. See HEC Electrical Box replacement document for details. Variation of a few units OK. |
| VFD BLW | Red or LCD | Reads 70 for non-wide and 60.1 for wide irrespective of state. Variation of 2 or 3 units in tenths place of display is OK. |
| LCD display | LCD | Please see Heat Exchanger Theory document for overview of all the menus provided by this display. |
Rear Pedestal
| Front Status Indicators | Status | Port Description | Indicator Description |
| +5D | Green | See Indicator Description. | Lit indicates +5V_dig from PDB/PDBv is OK. |
| -5V | Green | Lit indicates -5V from PDB/PDBv is OK. | |
| +5V | Green | Lit indicates +5V from PDB/PDBv is OK. | |
| +6V | Green | Lit indicates +6V from PDB/PDBv is OK. | |
| 1-4 | See Indicator Description. | P-conn. ch. 1-4, 16W16 | Following colors indicate to which IF receiver bank the 4 channels are being routed: Orange=Bank 1, Green=Bank 2, Blue=Bank 3, Violet=Bank 4, All Red=Loopback mode, All Violet=Reset in progress. Intended for engineering and research use. Rotating color display seen prior to successful TPS Reset. Little, if any, value to Service as these LEDs indicate internal RFSB switching. |
| 5-8 | P-conn. ch. 5-8, 16W16 | ||
| 9-12 | P-conn. ch. 9-12, 16W16 | ||
| 13-16 | P-conn. ch. 13-16, 16W16 |
| Front Status Indicators | Status | Port Description | Indicator Description |
| +5D | Green | See Indicator Description. | Lit indicates +5V_dig from PDB/PDBv is OK. |
| -5V | Green | Lit indicates -5V from PDB/PDBv is OK. | |
| +5V | Green | Lit indicates +5V from PDB/PDBv is OK. | |
| +6V | Green | Lit indicates +6V from PDB/PDBv is OK. | |
| B1 | See Indicator Description. | Body In 1, BNC | Following colors indicate to which IF receiver bank the 4 channels are being routed: Orange=Bank 1, Green=Bank 2, Blue=Bank 3, Violet=Bank 4, All Red=Loopback mode, All Violet=Reset in progress. Intended for engineering and research use. Rotating color display seen prior to successful TPS Reset. Little, if any, value to Service as these LEDs indicate internal RFSBv switching. |
| B2 | Body In 2, BNC (unused) | ||
| A1 | A-conn. 1-4, 8W8 | ||
| A2 | A-conn. 5-8, 8W8 | ||
| 4-Jan | P-conn. ch. 1-4, 16W16 | ||
| 8-May | P-conn. ch. 5-8, 16W16 | ||
| 12-Sep | P-conn. ch. 9-12, 16W16 | ||
| 13-16 | P-conn. ch. 13-16, 16W16 |
| Front Status Indicators | Status | Port Description | Indicator Description |
| +5D | Green | See Indicator Description. | Lit indicates +5V_dig from PDB/PDBv is OK. |
| -5V | Green | Lit indicates -5V from PDB/PDBv is OK. | |
| +5V | Green | Lit indicates +5V from PDB/PDBv is OK. | |
| +6V | Green | Lit indicates +6V from PDB/PDBv is OK. | |
| B1 | See Indicator Description. | Body In 1, BNC | Following colors indicate to which IF receiver bank the 4 channels are being routed: Orange=Bank 1, Green=Bank 2, Blue=Bank 3, Violet=Bank 4, All Red=Loopback mode, All Violet=Reset in progress. Intended for engineering and research use. Rotating color display seen prior to successful TPS Reset. Little, if any, value to Service as these LEDs indicate internal RFSB2 switching. |
| B2 | Body In 2, BNC (unused) | ||
| A1 | A-conn. 1-4, 8W8 | ||
| A2 | A-conn. 5-8, 8W8 | ||
| 4-Jan | P-conn. ch. 1-4, 16W16 | ||
| 8-May | P-conn. ch. 5-8, 16W16 | ||
| 12-Sep | P-conn. ch. 9-12, 16W16 | ||
| 13-16 | P-conn. ch. 13-16, 16W16 |
| Front Status Indicators | Status | Indicator Description |
| DRIVER 1 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 1 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 1 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 2 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 2 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 2 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 3 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 3 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 3 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 4 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 4 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 4 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 5 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 5 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 5 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 6 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 6 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 6 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 7 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 7 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 7 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 8 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 8 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 8 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 9 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 9 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 9 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| DRIVER 10 ENABLE | Dark | Green when Apex module enabled to source multicoil bias. |
| DRIVER 10 10V MODE | Dark or green | Green when Apex in 10V operating mode. |
| DRIVER 10 7V MODE | Dark or green | Green when Apex in 7V operating mode. |
| PWR GOOD +5V | Green | Lit indicates +5V from PDB/PDBv/ePDB is OK. |
| PWR GOOD -5V | Green | Lit indicates -5V from PDB/PDBv/ePDB is OK. |
| PWR GOOD -7V | Green | Lit indicates -7V from PDB/PDBv/ePDB is OK. |
| PWR GOOD +7V | Green | Lit indicates +7V from PDB/PDBv/ePDB is OK. |
| PWR GOOD +10V | Green | Lit indicates +10V from PDB/PDBv/ePDB is OK. |
| PWR GOOD -10V | Green | Lit indicates -10V from PDB/PDBv/ePDB is OK. |
| PWR GOOD +3.3V | Green | Lit indicates onboard regulated voltage is OK. |
| PWR GOOD +1.2V | Green | Lit indicates onboard regulated voltage is OK. |
| FAULTS BIAS | Dark or yellow | Yellow when system idle or when during active scanning bias current or voltage outside operating limits. |
| FAULTS TEMP | Dark | Yellow when board temperature exceeds operating limit. |
| Front Status Indicators | Status | Indicator Description |
| STRB | Green | Indicates strobe signal from RRx active. Dark when STRB fiber light missing at J13. |
| OSC off | Dark | Green when internal 40MHz digital core oscillator in standby mode during system scanning/receiving. |
| CAN err | Green | Flashing green in 2 ON, 2 OFF pattern after successful TPS Reset and during normal activity. Green alone, without CAN stat LED, during a disruption on CAN connection to CFB. |
| CAN stat | Green | Flashing green in 2 ON, 2 OFF pattern with CAN err LED during normal activity. |
| +5V | Green | Lit when +5V digital onboard power good. |
| +3.3V | Green | Lit when +3.3V digital onboard power good. |
| +1.2V | Green | Lit when +1.2V digital onboard power good. |
| FW_HB | Green | Flashing when firmware processor heartbeat OK. |
| +10V | Green | Lit when +10V from PDB/PDBv/ePDB OK. |
| -10V | Green | Lit when -10V from PDB/PDBv/ePDB OK. |
| +15V | Green | Lit when +15V from PDB/PDBv/ePDB OK. |
| -15V | Green | Lit when -15V from PDB/PDBv/ePDB OK. |
| IHBT | Dark or yellow | Yellow when inhibit signal active. Inhibit state often a part of normal operation and does not indicate a hardware malfunction. |
| UNBK | Dark | Green when unblank signal present. |
| LO-GD | Green | Lit when local oscillator signal good at J17. |
| LB-GD | Dark | Green when loopback signal present/active and good. |
| Front Status Indicators | Status | Indicator Description |
| +16V | Green | Power good status for voltage supplied by SRPS. |
| -16V | Green | Power good status for voltage supplied by SRPS. |
| +11V | Green | Power good status for voltage supplied by SRPS. |
| -11V | Green | Power good status for voltage supplied by SRPS. |
| +8V | Green | Power good status for voltage supplied by SRPS. |
| -8V | Green | Power good status for voltage supplied by SRPS. |
| +5V_CB | Green | Power good status for onboard voltage regulator. |
| +3V3 | Green | Power good status for onboard voltage regulator. |
| Front Status Indicators | Status | Indicator Description |
| +16V | Green | Power good status for voltage supplied by SRPS. |
| -16V | Green | Power good status for voltage supplied by SRPS. |
| +11V | Green | Power good status for voltage supplied by SRPS. |
| -11V | Green | Power good status for voltage supplied by SRPS. |
| +8V | Green | Power good status for voltage supplied by SRPS. |
| -8V | Green | Power good status for voltage supplied by SRPS. |
| +5V_CB | Green | Power good status for onboard voltage regulator. |
| +3V3 | Green | Power good status for onboard voltage regulator. |
| Front Status Indicators | Status | Indicator Description |
| FREQ_1 | Green or dark | Used with FREQ_2 to indicate field strength and selected nuclei. When green and FREQ_2 dark this indicates ePDB in 3T mode. Other combinations (dark, flashing) indicate 3T MNS nuclei selected. |
| FREQ_2 | Green or dark | Used with FREQ_1 to indicate field strength and selected nuclei. When green and FREQ_1 dark this indicates ePDB in 1.5T mode. Other combinations (flashing) indicate 3T MNS nuclei selected. |
| ANALOG_PW_ON | Green | Green when ePDB RFCB voltage regulators enabled. RFCB can enable and disable voltage regulators independant of ePDB. |
| 22V_IN | Green | Lit when +22V from eSRPS present. |
| TDM POWER GOOD | Green | Lit when the following voltages are normal: +6.85V_TDM, -6.85V_TDM, +5.05V_TDM, +5.05V_dig_TDM, +3.16V_dig_TDM. |
| HUB ANALOG POWER GOOD | Green | Lit when the following voltages are normal: +15V_RFCB, -15V_RFCB, +10V_RFCB, -10V_RFCB, +10V_MCDB1, +10V_MCDB2, -10V_MCDB1, -10V_MCDB2, +6V_RFSB, +5V_RFSB, -5V_ref. |
| HUB DIG POWER GOOD | Green | Lit when the following voltages are normal: +5V_dig_RFCB, +5V_dig_MCDB, +5V_dig_RFSB, +3.3V_dig_BKPL_RRx. |
| PDB INT POWER GOOD | Green | Lit when the following voltages internal to ePDB are normal: +15V_intermediate, +5V_intermediate, +3.3V_dig_ePDB. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| A-connector RED | Dark | Red when coil connected to ‘A’ and sensed, but coil ID invalid. |
| A-connector GREEN | Dark or green | Green when coil connected to ‘A’ and sensed and coil ID valid. |
| Home sensor LED | Dark or orange | Orange indicates table not at home, light beam unbroken. |
| Cradle detect LED | Dark or orange | Orange indicates cradle not connected, light beam unbroken. |
| EOT sensor LED | Dark or orange | Orange indicates table not at EOT, light beam unbroken. |
Magnet
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| HUB3V3 | Green | Indicates +3.3V power good from RF Hub PDB/PDBv/ePDB at RRx J1. |
| PORT ID2 | Dark or green | Green to indicate most significant bit 3 of RRx address from RRx J1. |
| PORT ID1 | Green to indicate bit 2 of RRx address from RRx J1. | |
| PORT ID0 | Green to indicate least significant bit 1 of RRx address from RRx J1. | |
| CLK ACTIVE | Green | Indicates 80MHz clock signal active from TDM and DTX1 at RRx J7. |
| PWR GOOD | Green | RRx power supply good. Power received from TDM at RRx J7. |
| IRF LINK UP | Green | Indicates IRF link at RRx J5 up. Flashing indicates IRF link error. |
| VRF LINK UP | Green | Indicates VRF link at RRx J4 up. Flashing indicates VRF link error. |
| Front Status Indicators | Status | Indicator Description |
| IRF link up 1 | Green | Indicates DVMR fiber optic link connection at J6 functional (lower 16 channels). |
| IRF link up 2 | Green | Indicates DVMR fiber optic link connection at J9 functional (upper 16 channels). |
| Inhibit | Dark | Yellow when inhibit signal active. |
| Inview | Dark | Green indicates FPGA inview active. |
| LO Active | Green | Indicates Local Oscillator signal present. |
| Spare | Dark | Green LED; not used. |
| VRF link up 1 | Dark | Green indicates VRF fiber optic link connection at J5 functional (lower 16 channels). Not used in systems with ICE. |
| VRF link up 2 | Dark | Green indicates VRF fiber optic link connection at J8 functional (upper 16 channels). Not used in systems with ICE. |
| Int done 1 | Green | Lower 16 channels RRx initialization OK. |
| Int done 2 | Green | Upper 16 channels RRx initialization OK. |
| Data overview 1 | Dark | Green indicates overflow in lower 16 channels receive. Not a fault. |
| Data overview 2 | Dark | Green indicates overflow in upper 16 channels receive. Not a fault. |
| TNS Notching 1 | Dark | Yellow indicates notching activity in lower 16 channels. |
| TNS Notching 2 | Dark | Yellow indicates notching activity in upper 16 channels. |
| RX blank | Dark | Green indicates all receive channels active. |
| CLK active | Green | 80MHz system clock received on J7 active. |
| RRx ID 2 | Dark or green | Green to indicate most significant bit 3 of RRx address from RRx J7. |
| RRx ID 1 | Dark or green | Green to indicate bit 2 of RRx address from RRx J7. |
| RRx ID 0 | Dark or green | Green to indicate least significant bit 1 of RRx address from RRx J7. |
| Power good | Green | Indicates power from mainboard regulators normal. |
| Front Status Indicators | Status | Indicator Description |
| FPGA Init Done | Green | Initialization OK. |
| CLK out Pres | Green | 80MHz clock signal received on J19 from DTX1 OK. |
| FPGA PWR Good | Green | Internal mainboard power OK. |
| Analog PWR Good | Green | Incoming 22V power OK. |
| Inhibit | Yellow | Indicates system inhibit signal active. |
| LO Detect | Dark | Green indicates Local Oscillator signal received on J10 from DTX1. |
| CH Overflow | Dark | Green when overflow detected in body receive signal. |
| VRF Linkup | Dark | Green indicates VRF fiber optic link connection at J15 active. Unused in systems with ICE. |
| Transient Detect | Dark | Yellow indicates active TDM notching. |
| Rx blank | Dark | Green indicates body receive channel inactive. |
| Inview | Dark | Yellow indicates FPGA inview active. |
| SSC Linkup | Green | Indicates DVMR fiber optic link connection at J16 functional. |
| Front Status Indicators | Status | Indicator Description |
| 1 | Green | Power on; Indicates onboard power supply functioning. |
| 2 | Green | NIOS heartbeat; Flashes show processor executing. |
| 3 | Green | 1W diag status; Indicates 1-wire network functioning. |
| 4 | Dark | Estop loop open; Red indicates E-Stop loop open. |
| 5 | Green | MCPDB CAN stat; Flashes 2 ON, 2 OFF pattern to indicate normal activity on MCPDB CAN connection. |
| 6 | Yellow | MCPDB CAN err; Flashing with LED 5 during normal operation or alone may indicate a fault on MCPDB CAN connection. |
| 7 | Green | SRI CAN stat; Flashes in 2 ON, 2 OFF pattern to indicate normal activity on CAN link. |
| 8 | Dark | SRI CAN err; Yellow indicates error or fault on CAN link. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
Patient Table
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
Magnet Monitor 3 (MM3)
| Front Status Indicators | Status | Indicator Description |
| Power | Green | MM3 power supply OK. |
| Heater | Yellow or dark | Yellow indicates when heater circuit active. |
| Alarm | Dark | Red indicates MM3 in alarm condition. |
Global Operator’s Console (GOC)
| Front Status Indicators | Status | Indicator Description |
| Unmarked on DVD drive | Dark or green | Flashes green to show drive read or write activity. |
| Unmarked on PC front | Green | Indicates PC power supply good. |
| Front Status Indicators | Status | Indicator Description |
| Power | Green | Lit to indicate normal power supply operation. |
| 1-8 or 1-16 | Green or yellow or dark | Solid green indicates connection at 1Gbit speed. Solid yellow indicates connection at 10/100Mbit speed. Flashing green indicates activity at 1Gbit speed. Flashing yellow indicates activity at 10/100Mbit speed. Dark indicates port unused or inactive. |
| Front Status Indicators | Status | Indicator Description |
| Tx | Dark or red | Flashes to indicate transmit activity from unit. |
| Rx | Dark or green | Flashes to indicate receive activity to unit. |
| PWR | Green | Lit when unit receiving power from Host USB. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| USB STATUS | Green | Indicates USB connections between IRD monitor and OCP trackballs. |
| LINK STATUS | Green | Indicates fiber connection between IRDM and IRD. |
| DVI STATUS | Green or orange | Green when DVI-D cable connected between Host, monitor port 2 and IRDM J2 and Host outputting video to IRDM. Orange when Host not outputting video to IRDM but DVI-D cable connected between Host and IRDM. Dark when DVI-D cable not connected between Host and IRDM. |
| POWER | Green | Indicates internal DC power supply functional. |
F-50 Shield Cooler Compressor
| Front Status Indicators | Status | Indicator Description |
| HIGH PRESSURE GAS BYPASS | Green | Indicates gas bypass in normal closed position. Dark indicates in open position. |
| OVER TEMP-ERROR He discharge | Dark | Red indicates helium gas discharged from compressor is over temperature. |
| OVER TEMP-ERROR He after cooler | Dark | Red indicates helium gas discharged from shield cooler is over temperature. |
| OVER TEMP-ERROR Water outlet | Dark | Red indicates water temperature discharged from compressor is over temperature. |
Operator’s Workspace
| Front Status Indicators | Status | Indicator Description |
| Unmarked | Blue or green | Blue Indicates monitor power displaying video; green indicates monitor not displaying video and in standby mode. |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| Channel 1 Alarm | Dark | Red during low O2 level condition. |
| Channel 1 Power | Green | Indicates unit powered and not in alarm condition. |
| Front Status Indicators | Status | Indicator Description |
| O2 % | Dark | At ISA-60M main module; red indicates low O2 condition. |
| Alarm 2 | Dark | At ISA-60M main module; red indicates low O2 condition. |
| POWER | Green | At ISA-60M main module; green indicates unit power normal. |
| FAULT | Dark | At ISA-60M main module; red indicates unit fault condition. |
| Power/Fault | Green | At remote MRI-5175 module; Green indicates unit powered on and in normal state. Red or flashing red and/or green indicates a fault condition or calibration state. |
| Alarm | Dark | At remote MRI-5175 module; red indicates low O2 condition. |
| Front Status Indicators | Status | Indicator Description |
| O2 ALARM | Dark | Flashing red during low O2 level condition or fault. |
| Power | Green | Indicates unit receiving AC power. |
| Unlabeled | Dark | At remote O2 sensor box; flashing red during low O2 condition. |
Multi-Nuclear Spectroscopy (MNS)
| Front Status Indicators | Status | Indicator Description |
| POWER | Green | System Controller power supply good. |
| STANDBY | Yellow | Indicates cabinet in STANDBY state. |
| READY | Dark | Green indicates System Controller in ready state. |
| UNBLANK | Dark | System Controller; Green indicates presence of unblank signal from BB RF AIF. |
| FAULT | Dark | System Controller; Red indicates fault condition. |
| STATUS CODE 5 | Dark or yellow | System Controller; yellow indicates most significant bit 6 of 6 bit status code. |
| STATUS CODE 4 | Dark or yellow | System Controller; yellow indicates bit 5 of 6 bit status code. |
| STATUS CODE 3 | Dark or yellow | System Controller; yellow indicates bit 4 of 6 bit status code. |
| STATUS CODE 2 | Dark or yellow | System Controller; yellow indicates bit 3 of 6 bit status code. |
| STATUS CODE 1 | Yellow or dark | System Controller; yellow indicates bit 2 of 6 bit status code. Lit alone means amp in transitional state. Bit 1 and this LED lit means amp in standby state. |
| STATUS CODE 0 | Yellow or dark | System Controller; yellow indicates least significant bit 1 of 6 bit status code. Lit alone means amp in operational state. |
| POWER | Green | At upper rear fan panel on cabinet are one pair of viewing holes for RF Preamp/Splitter module. LED visible through left hole indicates module power OK. |
| FAULT | Dark | At upper rear fan panel on cabinet are one pair of viewing holes for RF Preamp/Splitter module. Red LED visible through right hole indicates module fault. |
| POWER | Green | At rear fan panel on cabinet are one pair of viewing holes for each power module 1 (left) to 5 (right). LED visible through left hole indicates internal RF power module DC power OK. |
| FAULT | Dark | At rear fan panel on cabinet are one pair of viewing holes for each power module 1 (left) to 5 (right). Lit red LED visible through right hole indicates RF power module fault. |
| DIAGNOSTIC PS1 | Dark | See from rear of power supply module. Red indicates bad incoming AC power, failed fan, obstructed air flow, or failed power supply module. |
| DIAGNOSTIC PS2 | Dark | See from rear of power supply module. Red indicates bad incoming AC power, failed fan, obstructed air flow, or failed power supply module. |
| Front Status Indicators | Status | Indicator Description |
| POWER | Green | Indicates DC power supply inside chassis module is OK. |
| FAULT | Dark | Red indicates fault inside chassis module. |
| READY | Green | Indicates unit in operational state. |
| UNBLANK | Dark | Green indicates presence of active unblank signal from cabinet J1. |
| STANDBY | Yellow | Indicates unit in STANDBY state. |
| STATUS CODE 5 | Dark or yellow | Driver Chassis; yellow is most significant bit 6 of 6 bit status code. |
| STATUS CODE 4 | Dark or yellow | Driver Chassis; yellow is bit 5 of 6 bit status code. |
| STATUS CODE 3 | Dark or yellow | Driver Chassis; yellow is bit 4 of 6 bit status code. |
| STATUS CODE 2 | Dark or yellow | Driver Chassis; yellow is bit 3 of 6 bit status code. |
| STATUS CODE 1 | Yellow or dark | Driver Chassis; yellow is bit 2 of 6 bit status code. Lit alone amp in transitional state. Bit 1 and this LED lit amp in standby state. |
| STATUS CODE 0 | Yellow or dark | Driver Chassis; yellow is least significant bit 1 of 6 bit status code. Lit alone amp in operational state. |
| AC ON | Green | Indicates AC power present at power supply chassis. |
| AC FAULT | Dark | Red indicates fault internal to power supply chassis. |
MRU
| Front Status Indicators | Status | Indicator Description |
| HEATER TEST | Dark | Case 1 - Solid green indicates LED operation normal in response to pressing TEST HEATER LED button with TEST HEATER toggle switch in center location. Case 2 - Solid green indicates heater circuit ‘A’ or ‘B’ continuity normal in response to moving TEST HEATER toggle switch to ‘A’ or ‘B’ locations while TEST HEATER LED button is not pressed. |
| CHARGER POWER | Green | Indicates unit receiving AC power and battery charger operating normally. |
| BATTERY TEST | Dark | Green indicates enough charge for battery to illuminate LED in response to pressing TEST BATTERY button. |
| HEATER ACTUATED | Dark | Red indicates heater circuit actuated in response to pressing red RUN DOWN button. This LED may also sometimes falsely illuminate during routine testing. Internal reset of unit required to extinguish LED. |
MDP
| Front Status Indicators | Status | Indicator Description |
| POWER ON | Green | Indicates MAIN INPUT CB1 ON and incoming site AC power present. |
PET/MR
| Front Status Indicators | Status | Indicator Description |
| Leak Sensor LED | Green | Green when sensor not faulted and dry. Red when sensor faulted or wet. |
| Front Status Indicators | Status | Indicator Description |
| Left LED above J15 | Green | Lit indicates 80V1 from 80V PS normal. |
| Inner-left LED above J15 | Lit indicates HV1 and LV1 voltages from DMOD Low Voltage PS normal. | |
| Inner-right LED above J15 | Lit indicates HV1 from DMOD Low Voltage PS normal. | |
| Right LED above J15 | Dark | Green indicates clock signal present at CPDA J20 from PROP. If present this clock signal supplies J8 to J14. |
| Left LED above J19 | Green indicates clock signal present at CPDA J20 from PROP. If present this clock signal supplies J1 to J7 and clock signal output at J21. | |
| Inner-left LED above J19 | Green | Lit indicates HV2 from DMOD Low Voltage PS normal. |
| Inner-right LED above J19 | Lit indicates HV2 and LV2 voltages from DMOD Low Voltage PS normal. | |
| Right LED above J19 | Lit indicates 80V2 from 80V PS normal. |
| Front Status Indicators | Status | Indicator Description |
| QUAD 1, 2, 3, 4 DC OK | Green | Lit when internal Quad 1, 2, 3, and 4 DC supply voltages normal and Output Enable switch in up position. |
| Front Status Indicators | Status | Indicator Description |
| Above QUAD 1 | Green | |
| Above QUAD 2 | Lit when internal 80V power supply voltages supplying Quad 1, 2, 3, and 4 normal. | |
| Above QUAD 3 | ||
| Above QUAD 4 |
| Front Status Indicators | Status | Indicator Description |
| None | None | None |
| Front Status Indicators | Status | Indicator Description |
| PWR FAIL | Dark | Red indicates one or more power supplies failed. |
| WDOG | Green | Flashing indicating watchdog process running. |
| FLASH | Dark | Flashing yellow indicates code flash activity. |
| RST | Dark | Reset; Green during reset activity. |
| LD | Dark | Loading; Flashing green during code flashing activity. |
| SAF | Dark | Safe; Yellow indicates operating in default safe mode. |
| USR | Green | User; operating in user mode. |
| ERR | Dark | Error; Red indicates error detected during flash download. |
| Up and down arrows under/over fibers | Green | Indicates each Tx and Rx section of under or over fiber connected and active. |
| Front Status Indicators | Status | Indicator Description |
| Power | Green | Lit to indicate normal power supply operation. |
| 1-16 | Green or yellow or dark | Solid green indicates connection at 1Gbit speed. Solid yellow indicates connection at 10/100Mbit speed. Flashing green indicates activity at 1Gbit speed. Flashing yellow indicates activity at 10/100Mbit speed. Dark indicates port unused or inactive. |
| Front Status Indicators | Status | Indicator Description |
| Power switch | Blue | Power on. |
| Hard drive indicators | Green or orange | Green but flash orange during hard drive access. |
| Front Status Indicators | Status | Indicator Description |
| LCD | Many | Numerous menus & levels. See vendor chiller manual under Installation & Upgrade folder, PET-MR folder on service methods. |
| Front Status Indicators | Status | Indicator Description |
| ɸA | Green (Indicators may prematurely fail and go permanently dark. See *NOTE: comments below) | Lit when incoming AC power on phase A and main 15 kVA PDU circuit breaker closed/engaged. See *NOTE: comments below. |
| ɸB | Lit when incoming AC power on phase B and main 15 kVA PDU circuit breaker closed/engaged. See *NOTE: comments below. | |
| ɸC | Lit when incoming AC power on phase C and main 15 kVA PDU circuit breaker closed/engaged. See *NOTE: comments below. | |
| ESTOP ON | Lit in normal operation after PROP goes active. Dark during E-Stop event or anytime PROP not active. See *NOTE: comments below. |
These indicators utilize neon lamps. It has been observed that neon lamp life-span can be short with neon lamps failing to illuminate after 1 year of use. When this happens, the indicators will be dark even though the circuits are active. Do not rely on status of indicators alone when assessing state of PET Cabinet PDU.
Glossary
| Acronym | Description |
| 8W8 | Refers to 8 coaxial connections within a single sub-D connector. |
| 16W16 | Refers to 16 coaxial connectors within a single sub-D connector. |
| A1 | A-connector channels 1 to 4 |
| A2 | A-connector channels 5 to 8 |
| AC | Alternating Current |
| ACT | ACTivity |
| ASC | Amplifier Support Controller |
| B1 | Body receive 1 |
| B2 | Body receive 2 (may not be used) |
| BLW | BLoWer |
| BNC | Bayonet Neill-Concelman connector |
| CAM | Combined Asc Mgd |
| CAN | Controller Area Network |
| CH | CHannel |
| CLK | CLocK |
| COMM | Communication |
| CPU | Central Processing Unit; describes a semiconductor processing chip. |
| DCHV | Direct Current High Voltage |
| DD | Dynamic Disable and Direct Drive |
| DMOD | DeMODulator |
| DSP | Digital Signal Processing |
| DTX | Digital Transmit |
| DUART | Dual Universal Asynchronous Receiver Transmitter |
| DVD | Digital Video Disk |
| DVI | Digital Video Interface |
| DVMR | Dynamic Volumetric Magnetic Resonance |
| E Stop | Emergency Stop |
| EOT | End Of Travel |
| ERR | ERRor |
| eSRPS | eco-Scan Room Power Supply |
| FLT | FauLT |
| FPGA | Field Programmable Gate Array; this is a semiconductor chip. |
| FW | FirmWare |
| GD | GrounD |
| GMC | Gradient Master Controller |
| GPMP | Gradient PuMP |
| HB | HeartBeat; usually shows that a processor is running. |
| HS | Hot Swap; this method of replacement is not supported. |
| HSS | High Speed Stability test |
| HV | High Voltage |
| ICEFL | Integrated Controller Engine FauLt |
| ICN | Image Compute Node |
| ID | Identification |
| IDB | Integrated Daughter Board |
| IGBT | Insulated Gate Bipolar Transistor |
| IHBT | InHiBiT |
| IRF | Integrated functions Remote Fiber optic link |
| kVA | Kilo-Volt Amperes |
| LB | LoopBack signal |
| LCD | Liquid Crystal Display |
| LD | LoaD |
| LED | Light Emitting Diode |
| LL0L | Low voltage Left 0 (zero) Lower |
| LO | Local Oscillator signal |
| LPCA | Low Profile Carriage Assembly |
| LR0U | Low voltage Right 0 (zero) Upper |
| LV0 | Low Voltage 0 (zero) |
| MCPDB | Motor Control Power Distribution Board (located inside PHPS2) |
| MDP | Main Disconnect Panel |
| MGD | Multi-Generational Data acquisition |
| MNS | Multi-Nuclear Spectroscopy |
| NIOS | Describes a type of FPGA processor. |
| O2 | Oxygen |
| OCT | Octavius |
| OSC | OSCillator signal |
| PC | Personal Computer |
| PDB | Power Distribution Board |
| PET | Positron Emission Tomography |
| PMON | Power MONitor |
| PPMP | Pgr cabinet PuMP |
| PS | Power Supply |
| PWR | Power |
| QUAD | Describes a quantity of 4. |
| RDY | ReaDY |
| REF | REFerence; sometimes refers to a clock signal. |
| RF | Radio Frequency |
| RFLK | RF Lock |
| RRx | Remote Receiver |
| RST | ReSeT |
| RX | Receive |
| S1 | Status 1; indicates an arbitrarily assigned software state. |
| S2 | Status 2; indicates an arbitrarily assigned software state. |
| S3 | Status 3; indicates an arbitrarily assigned software state. |
| SAF | SAFe |
| SCIM | Scan Control Interface Module |
| SFP | Small Form-factor Pluggable |
| SPU | Serial Processing Unit |
| SRPS | Scan Room Power Supply |
| SSC | System Sequence Control; can refer to DVMR Link or to component (ICE or CAM) |
| STAT | STATus |
| STRB | STRoBe signal |
| TDM | Transient Detection Module |
| TNS | Transient Noise Suppressor |
| TPS | Tardis Processing System |
| TR | Transmit / Receive |
| TX | Transmit |
| UNBL | UNBLank |
| USB | Universal Serial Bus |
| USR | USeR |
| V | Voltage |
| VFD | Variable Frequency Drive |
| VRF | Volume Recon Fiber optic link |
| WARP | WAveform Rotation Processor |
| WDOG | Watch DOG; describes a timing process. |
| XGA | eXtreme Gradient Amplifier |
| XGD | eXtreme Gradient Driver; refers to entire gradient subsystem. |
| XPS | eXtreme Power Supply |