Body Coil Mount and Base Plate (Front) Replacement
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 2 | Not Applicable | 2 hours | 60 minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Nonmagnetic Titanium Service Tool Kit, Large Set | 1 | - | 5112581 | - |
| Cut-Resistant Gloves | 1 | - |
| - |
| non-magnetic small level | 1 | - |
| - |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Front Body Coil Mount | 1 | - |
Refer to FRU manual | - |
| ||||
About this task
Overview
Use this procedure to replace the body coil mount and base plate (front).
Body Coil Mount Front Assembly Removal
Procedure
Front Base Plate Replacement
Procedure
Restoration
Procedure
Finalization
Procedure
- Remove LOTO. Refer to Removing LOTO - ISC.
- Move the table up and restore the dock interface cover.
- Perform Dual Drive Quadrature Calibration. Refer to Dual drive quadrature tool.
- Perform the system performance test (SPT).