• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA3002D230GYZ
  • Topic ID: id_15667342 Version: 1.12
  • Date: Jan 26, 2022 4:07:28 PM

VRMw shim retaining ring segment replacement

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-6 hours-
Tools and test equipment
ItemQuantityPart numberManufacturer
VRMw Shim Retaining Ring FRU (contains 6 screws and pack of sandpaper)15542958-
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
B0 Power Supply Kit12141701-
STET tool1

5431827

-
Consumables
ItemQuantityPart numberManufacturer
SandpaperAs needed--
Required conditions

Remove patient front end bell and patient support bridge from the magnet bore as described in appropriate procedures before starting work.

STET assembled and installed according to STET Operator Manual (5431810).

Table 1. Safety

About this task

This procedure describes how to replace the shim retaining ring on the VRMw gradient coil while the magnet is ramped. The retaining ring is split into 3 segments, which can be replaced individually. The segments are split between shim trays 36 and 0, trays 12 and 13, and trays 24 and 25.

Figure 1. Shim retaining ring

Procedure

  1. Note: Induced current must be removed from the B0 coil before shimming the magnet. The B0 coil power supply kit is used to quench the induced currents in the B0 coil circuit. The B0 coil power supply is switchable from 115/230 VAC, 50/60 Hz. The B0 coil current must be removed whenever a shim drawer has been moved during the shimming process.
    Remove the passive shim trays from the gradient coil in the section where the ring segment is to be replaced.

    Refer to the STET manual (5431810) for the shim tray removal process.

  2. Using non-ferrous Phillips screwdriver and non-ferrous channel lock pliers, remove the 6 stainless mounting screws on the segment of the ring to be replaced.

    You may need to tap on the screw heads to break the Loctite bond before you are able to remove them.

    Figure 2. Removing stainless mounting screws
  3. Pry loose and take out the broken ring segment. Start prying from the end of the segment to help break the epoxy bond.
  4. Using sandpaper, clean the mounting surfaces on the gradient coil of any resin fragments to ensure a good fit of the ring segment (especially the mounting surface on the inner surface of the outer gradient coil).
  5. Replace the broken ring segment with the new one.
    Figure 3. VRMw gradient coil shim retaining ring
  6. Secure each ring segment to the coil with (11 off) thread cutting screws (2162974-13) and (2 off) thread cutting screws (2162974-12) at both end of the ring.
    Figure 4. Secure each ring segment
  7. Safely re-install all passive shim trays. Be sure to return each tray to the correct slot to avoid changing the shim of the magnet.

Finalization

  1. De-install STET as described in the STET Operator Manual (5431810).
  2. Re-install the bridge and front end bell.
  3. Connect the B0 coil power supply output cable to P304-1 on the magnet connector box.
  4. Connect the B0 coil power supply input cable between the power supply and the voltage source.
  5. Press the B0 coil power supply switch (the switch is spring-loaded) and hold for one minute.

    The LED lights to indicate the power supply is supplying output current.

  6. Wait for three minutes for the magnet to recover before performing LVShim.
  7. Do an LVShim scan (see LVShim Procedures).
  8. Do a Check Scan.