• Topic ID: id_15271093
  • Version: 3.0
  • Date: Aug 29, 2019 1:44:05 AM

Dual Table Alignment

Prerequisites

Table 1 Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 2 hours -
Table 2 Tools and test equipment
Item Quantity Effectivity Part number Manufacturer
Non-ferrous medium slotted head screwdriver 1 - - -
Non-ferrous medium Phillips head screwdriver 1 - - -
Table 3 Consumables
Item Quantity Effectivity Part number Manufacturer
Marking tape 12 in (305 mm) - - -
Table 4 Safety
Notice Image
  • warning
  • STRONG MAGNETIC FIELD!
  • FERROUS HAND TOOLS CAN BECOME DANGEROUS PROJECTILES IN THE VICINITY OF THE SYSTEM MAGNET.
  • Do not use any magnetic/ferrous tools or equipment inside the magnet room.

Overview

This procedure is for the alignment of two tables dedicated to one MR suite. You may need to adjust all four wheels on both tables.

Procedure

  1. Drive the LPCA through the magnet to the end of the rear pedestal.
  2. Undock the patient table and move to an open work area.
  3. Remove all lower table covers from all patient tables adjusted in this procedure (see Patient Table Upper Bellows and Lower Base Side Cover Removal).
  4. Confirm the following:
    1. Bridge height is 502–505 mm.
    2. Bridge is centered left/right and in/out.
    3. Bridge is level (left/right, front/back).
  5. Notice Image
    • warning
    • Personnel INJURY or equipment damage!
    • The Dock Contains Ferrous Materials and may be pulled into the magnet bore if securing hardware is removed.
    • Only loosen the hardware sufficiently to perform adjustments in this procedure. ensure THE DOCK remains secured TO PREVENT IT FROM BEING PULLED INTO THE MAGNET BORE.
  6. At the rear of the dock, loosen the top and bottom dock bracket nuts (2 on each side).

    Figure 1. Dock Bracket Nuts

  7. Get the table removed earlier (first table) and check table alignment/levelness with the bridge to ensure smooth travel. Adjust the height as needed according to Leveling Patient Table.
  8. Adjust left/right according to Dock Centering and Dock Hook Tension Adjustment).
  9. Undock the first table and get the second table.
  10. Dock the second table and repeat Step 6 and dock hook centering procedure from Step 7, but do not perform dock centering.
  11. Mark the proper dock alignment (left/right) of the second table with a second piece of tape.
    1. Mark the center point between the two marks.
    2. Shift the dock to align with the center point mark.

      Figure 2. Marking Center Point

  12. Secure the dock by tightening the dock bracket nuts at the rear of the dock. See Figure 1.

Finalization

  1. Dock each table, move the cradle into the bore, and confirm smooth table travel.
  2. If necessary, adjust the table height (see Table Top Height Adjustment).