- SIGNA™ Hero 3.0T Service Methods
- 5852800-8EN Revision 1.0
- 00000018WIA30672110GYZ
- id_20035177.5
- Jul 7, 2021 3:51:57 PM
Removing the rear endbell
Removes the rear endbell from the magnet enclosure.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 1 hour and 5 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
- Applying LOTO - ISC
- Removing the magnet rear trim ring
- Removing the magnet rear cable cover
- Removing the rear bridge cover
- Removing the rear bridge inner cover
- Removing the rear endbell