Installing the DS Digitized Per Pin (DPP) Receiver Distributor (Rx Dist)
Install a DS Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) in the magnet outer electronics.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
15 minutes
-
Tools and test equipment
Item
Quantity
Part number
Manufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set
1 Kit
5113258
-
Nonmagnetic Titanium Service Tool Kit, Large Set
5112581
-
Replacement parts
Item
Quantity
Part number
Manufacturer
Rx Distribution FRU
1
5434906
-
Safety
CAUTION
Strong magnetic field
The DS Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) has no ferrous material and will not be attracted to the magnet, but the eddy currents can cause resistance when you pull the DS DPP Rx Dist away from the magnet.
When removing and replacing the DS DPP Rx Dist, keep the receiver as far away from the magnet as possible while moving it.
Procedure
Align the two keyhole shaped screw holes and two slotted screw holes on the DS DPP Rx Dist over the screws on the side electronics plate.
Make sure the blank side of the DS DPP Rx Dist module points to the right.