Installing the blower box filter
Procedure
- Install the EMC filter to the blower box housing.
- Install the charcoal air inlet filter(s).
- Install and hand-tighten each nut.
- Tighten each nut securely using a non-ferrous 7 mm wrench.
Figure 1. Blower box filter - patient air side