• Topic ID: id_2011632
  • Version: 4.0
  • Date: Jul 19, 2019 10:54:58 AM

Installing the blower box filter

Procedure

  1. Install the EMC filter to the blower box housing.
  2. Install the charcoal air inlet filter(s).
  3. Install and hand-tighten each nut.
  4. Tighten each nut securely using a non-ferrous 7 mm wrench.

    Figure 1. Blower box filter - patient air side