• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WIA30324630GYZ
  • Topic ID: id_2011314 Version: 7.3
  • Date: Jul 5, 2021 10:27:23 PM

Replacing the ICN - Dell R640

Remove and install the Image Compute Node (ICN).

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
110 minutes40 minutes5 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Replacement parts
ItemQuantityPart numberManufacturer
Dell R640 (Gen7-DL) Performance ICN with Dolphin Interface Card and GPU Card (For software version DV29.1 and later) 15941000-9-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

Before doing this replacement, you must shut down the system.

Procedure

  1. From the header area of the screen, click the Tools icon.
  2. Click the Service Desktop Manager tab, and then click Utilities.
  3. Select System Shutdown.
  4. Make sure the computer is turned off.