Removing the Integrated Control Engine (ICE) fan assembly

Removing the Integrated Control Engine (ICE) fan assembly from the System Cabinet.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 to 10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set1 Kit5113258-
Nonmagnetic Titanium Service Tool Kit, Large Set5112581-

Procedure

  1. From the Common Service Desktop, turn off the ICN(s) (see ICN On/Off Procedure).
  2. Do LOTO for reconstruction equipment. See the latest revision of the MR Service Safety Manual (5452735), available from the online documentation library.
  3. Open the PGR door.
  4. Remove the ICE as follows:
    Note: Do not remove the slide rails from the ICE chassis.
  5. Press the release tabs in on each side and pull the ICE forward.
    Note: Hold the ICE carefully when you remove it so that it does not fall.
  6. Do the following steps to remove a fan from the back of the ICE.
    1. Pull the fan power cable from the hole beside the fan, and disconnect the cable.
      Note: Be careful when you remove the power cable. It can fall into the hole shown below.
      Figure 1. Hole for power cable
    2. Remove four screws, and remove the fan.
    Figure 2. Removing the fan
    1Connect the fan cable to the power cable
    2Tighten four screws