Installing the Waveguide sub-panel

Prerequisites

The original waveguide sub-panel must be removed and replaced with the custom waveguide to allow cabling into the magnet room.

  • danger
  • Fatal electric shock hazard
  • Lethal voltages are present in the PGR Power Distribution Unit (PDU) or gradient cabinet.
  • Perform a system shutdown, and then perform LOTO on the main PDU breaker and the penetration cabinet. The capacitors may have 800 volts. Wait at least five minutes after LOTO for the capacitors to drain before continuing this procedure.
  • danger
  • Fatal Electric Shock Hazard
  • Coldhead (cryocooler compressor) presents a shock hazard.
  • Turn off and perform LOTO on the coldhead before continuing this procedure. See the MR Service Safety Manual (5452735).

Procedure

  1. Perform LOTO on the PGR and penetration cabinets .
  2. Perform LOTO on the Heat Exchanger Cabinet (HEC) to remove power from the cryocooler compressor, which is necessary to complete this procedure.
  3. Install the custom waveguide panel (2394034-10) in place of the standard panel.

    Figure 1. Sub-panel

    note: If the waveguide panel has preexisting (custom) modifications, they must be reapplied to the sub-panel before BrainWave can be installed.