• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WIA304C5C40GYZ
  • Topic ID: id_2019263 Version: 3.0
  • Date: Apr 29, 2020 2:59:35 PM

Raising the dockable table bellows

Detaches the dockable table bellows cover from the table base covers and raises the bellows.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Remove the screws securing the bellows cover to the table base covers.
    Figure 1. Bellows lower screws
  2. Lift the bellows cover and secure it in the raised position with the Velcro strips.