- SIGNA™ Hero 3.0T Service Methods
- 5852800-8EN Revision 1.0
- 00000018WIA300D4840GYZ
- id_20149134.1
- Jul 12, 2021 1:02:57 PM
Replacing the GOC hard drive - Dell T5820
Removal and installation instructions for replacing the solid state drive (SSD) in the host computer.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | 2 hours |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Screwdriver | 1 | - | - |
| Nonmagnetic Titanium Service Tool Kit, Small Set (can be used on magnets with magnetic fields > 1.5T) | 1 | 5113258 | - |
| Disk Management Tool | 1 | - | - |
| ESD Mat | 1 | - | - |
| ESD Strap | 1 | - | - |
| Brass Master Padlock with Brass Shackle | 2 | 46-194427P320 | - |
| Red LOTO Personal Lock Wrap | 1 | 2393068 | - |
| Multi-Locking Device (if multiple service personnel are involved) | 1 | 46-194427P313 | - |
| Red Warning LOTO Tag | 1 | 46-194427P322 | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| GE Equipment Maintenance Sticker | 1 | 5661793 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| GOC Dust Filter for T5820 | 1 (if needed) | 5793940 | - |
| Adhesive-Backed Velcro, 2 x 4 inches | 2 (if needed) | 46-317475P2 | - |
| 512GB SATA III SFF SSD 3D-TLS or V NAND Memory | 2 | 8780011-102 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |