• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA303FB230GYZ
  • Topic ID: id_15667179 Version: 2.1
  • Date: Jul 13, 2021 4:27:37 PM

Replacing the Global Operator Cabinet Audio Assembly (GOCAA)

Removal and installation instructions for replacing the Global Operator Cabinet Audio Assembly (GOCAA).

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes30 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Screwdriver1--
Replacement parts
ItemQuantityPart numberManufacturer
GOC Audio Assembly12395043-2-
Required conditions

The host computer and the GOC audio assembly must be powered off.

Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Shut down the application software on the Host computer.
  2. Lock out and tag out power sources. See Lockout/Tagout of the GOC.