• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA305DC230GYZ
  • Topic ID: id_15667322 Version: 2.1
  • Date: Jul 13, 2021 4:27:36 PM

Replacing the temperature and humidity sensor

Removes the temperature and humidity sensor from the Global Operator Cabinet (GOC) and installs a replacement sensor.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 minutes-
Table 1. Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Table 2. Replacement parts
Replacement parts
ItemQuantityPart NumberManufacture
Temperature & Humidity Sensors Kit 15147136-2-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.