The 1.5T GEM Coil MCQA will use the 1.5T phantoms, which are green in color.
Note:
Coils do not ship with phantoms. Phantoms come in a Unified Phantom set with the MR System.
Table 2. Tools and test equipment
Item
Quantity
Effectivity
Part number
Manufacturer
TL Unified Phantom (for 1.5T)
2
-
5343347
-
PA Phantom Positioner (for both 1.5T and 3.0T)
1
-
5405748
-
Procedure
Place the flat filler panel on the GEM table in the recess closest to the magnet bore.
Figure 1. Flat filler panel
1
Flat filler panel
2
Bore
Landmark at the straight edge of the flat filler panel.
Figure 2. Landmark location
1
Landmark/bottom edge of the flat filler panel
2
Bore
From Landmark 0 mm (at the bottom edge of the flat filler panel), move the table from bore until the system LCD displays s 50 mm.
Figure 3. Phantom position
1
Bore
2
s 50 mm
3
50mm overlap
4
Landmark 0 mm
Place two TL unified phantoms on the GEM table centered in the right to left direction and the front edge of the phantom closest to the bore aligned with the s 50 mm laser mark.
Important: Make sure that the front end of the first phantom (closest to the bore) overlaps the bottom edge of the flat filler panel 50mm. A slight misalignment of the phantoms and landmark may cause a false negative result from MCQA.
Important: Do not change the landmark.
Press Advance to Scan.
Run the MCQA Tool. Make sure that the correct coil is selected in the Current Coil field:
For the original PA coil: 1.5T_GEM_PA1
For the latest PA coil: 1.5T_GEM_PA_1 (note the extra underscore between "PA" and the number)