• Object ID: 00000018WIA30F8B030GYZ
  • Topic ID: id_12373976 Version: 1.5
  • Date: Sep 2, 2020 12:56:33 PM

GEM Coil Troubleshooting

Basic GEM PA Coil Theory

For systems with an RF Hub, multicoil bias is supplied on each of the receive lines from each of the RF Switch Board P1 and P2 connectors. This bias is used to enable and disable the various PA coil element receive arrays. The MC bias flows from the RF Switch Boards and through the MUX Module. The MC bias is not switched inside the MUX Module. The MC bias is bypassed around the switching circuitry inside the MUX Module and passes through the J1-PA, J2-PA, and J3-PA connections directly to the PA Coil. Likewise, with the exception of MUX select bits 1-4 which control the MUX switching array and +10VDC Auxiliary Power lines which supply power to the MUX Module switching array, the remaining Coil ID, TX Enable, and Coil Present lines from the RF Control Board pass directly through the MUX Module and to the PA Coil through the J4-PA connection.

RF receive signals from the patient or phantom are detected by the selected PA Coil elements and then routed through the J1-PA, J2-PA, and J3-PA connections to the MUX Module. RF switching circuits inside the MUX Module route the signals onto the proper receive channels to the RF Switch Boards inside the RF Hub. The configuration of the RF switching circuits inside the MUX Module is controlled by the binary setting of the MUX Select bits 1-4.

Figure 1. GEM Table Receive Chain (RRX) Block Diagram

PA Coil Troubleshooting Hints and Tips

  1. Use Coil Datapath Diagnostic (CDD), which uses MCRv Tool, to troubleshoot PA coil multi-coil bias and coil signal problems. MCRv Tool can be attached to cabling that normally connects to MUX Module inside patient table. See Figure 2. This setup and usage of MCRv Tool is discussed in more detail in CDD service methods document.

  2. Figure 1 shows interconnections between PA coil and RFSB and RFCB inside RF Hub and how it can be used to check continuity of cables, however, this is often done more quickly and thoroughly by using CDD.

  3. Troubleshoot P3(PA) multi-coil bias short-circuit faults by doing these things:

    1. Inspect electrical P3 (PA) connector on patient table and dock for damaged pins.

    2. Remove receive line where fault reported or all lines and observe that short- circuit fault is no longer reported. Repeat at every junction in chain until fault isolated.

    3. Use CDD and MCRv to help isolate fault location. See CDD service methods document.

  4. Troubleshoot P3(PA) multi-coil bias open-circuit faults by doing these things:

    1. Inspect electrical P3(PA) connector and patient table and dock for damaged pins.

    2. Open cradle and reseat 3 PA coil cable connections and watch for recessed/pushed receive line connections.

    3. Use CDD and MCRv to help isolate fault location. See CDD service methods document.

  5. PA Filler assembly can be installed in place of PA coil during troubleshooting and to permit customer scanning using coils other than PA coil. PA Filler contains only a Coil ID chip but does not contain any coil elements and does not require or use multi-coil bias.

  6. Multi-coil bias passes around the switching circuitry inside the MUX Module. It is not switched inside the MUX Module. It therefore stands to reason that multi-coil bias failures are unlikely to be due to MUX Module. These failures are more likely result of connector, cabling or PA coil issues.

Using MCRv (Coil Datapath Diagnostic)

The MCRv Tool can be used to verify that the Receive Chain is functioning correctly from the PA Mux Box output to the RFSB in the Rear Hub. The PA Mux Box and path up to the PA Coil is NOT tested. Cables at J1, J2, and J3 on MUX Module can be disconnected and reattached to MCRv Tool for testing. Coil Datapath Diagnostic (CDD) can then be used to test all signal paths from MCRv Tool back to RF Hub. This eliminates manual "signal chasing" using a DVM and probes. Figure 2 is from CDD service methods document and shows setup using MCRv Tool.

Figure 2. MCRv Tool connection at Mux Box

GEM PA Coil Signal Path to RF HUB

Table 1 represents the signal paths from the GEM PA Coil back to the RFSB Board. The table should be used as a reference — the Coil Datapath Diagnostic should be used from Mux back to the RFSB.

This table is also useful for running continuity checks between the PA Coil connectors to the Output of the Mux Box. To do this:

  1. Disconnect the Cables from the PA Coil at the PA Coil side.

  2. Disconnect the Mux Output Cables from the Mux (J1 and J2).

  3. Use a DVM and run a continuity check between the Mux Box and PA coil cables in the cable take-up.

Note:

PCX Connectors are internal to Mux. These are listed as Reference only. MUX Box is not intended to be serviced internally.

Table 1. Signals Path from GEM PA Coil to RFSB Board
RF Channel DC Rear Hub DOCK MUX PA Coil
RFSB Pin# Connector Pin# Connector PCX Connector Connector J1-PA Element 3T Element 1.5T
19 DC19 Slot 1-P3 11 P3 G3 J2 - 3 J8 1 L38 L37
23 DC23 Slot 2-P3 11 P3 K3 J2 - 7 J10 2 L36 L35
31 DC31 Slot 2-P3 15 P3 L3 J2 - 15 J17 3 L34 L33
27 DC27 Slot 1-P3 15 P3 H3 J2 - 11 J15 4 L32 L31
N/A None None N/A N/A N/A N/A J42 5 S2 S2
18 DC18 Slot 1-P3 10 P3 G2 J2 - 2 J12 6 L21 L19
22 DC22 Slot 2-P3 10 P3 K2 J2 - 6 J14 7 L18 L16
30 DC30 Slot 2-P3 14 P3 L2 J2 - 14 J32 8 L15 L13
3 DC3 Slot 1-P3 3 P3 E3 J1 - 3 J28 9 L30 L29
7 DC7 Slot 2-P3 3 P3 I3 J1 - 7 J30 10 L28 L27
15 DC15 Slot 2-P3 7 P3 J3 J1 - 15 J41 11 L26 L25
11 DC11 Slot 1-P3 7 P3 F3 J1 - 11 J31 12 L24 L23
26 DC26 Slot 1-P3 14 P3 H2 J2 - 10 J13 13 L12 L10
2 DC2 Slot 1-P3 2 P3 E2 J1 - 2 J30 14 L9 L7
6 DC6 Slot 2-P3 2 P3 I2 J1 - 6 J29 15 L6 L4
14 DC14 Slot 2-P3 6 P3 J2 J1 - 14 J40 16 L3 L1
J2-PA
N/A None None N/A N/A N/A N/A J44 1 L19 L21
N/A None None N/A N/A N/A N/A J35 2 L16 L18
N/A None None N/A N/A N/A N/A J34 3 L13 L15
N/A None None N/A N/A N/A N/A J16 4 L10 L12
20 DC20 Slot 1-P3 12 P3 G4 J2 - 4 J11 5 L37 L38
24 DC24 Slot 2-P3 12 P3 K4 J2 - 8 J10 6 L35 L36
32 DC32 Slot 2-P3 16 P3 L4 J2 - 16 J19 7 L33 L34
28 DC28 Slot 1-P3 16 P3 H4 J2 - 12 J18 8 L31 L32
N/A None None N/A N/A N/A N/A J21 9 L7 L9
N/A None None N/A N/A N/A N/A J9 10 L4 L6
N/A None None N/A N/A N/A N/A J20 11 L1 L3
10 DC10 Slot 1-P3 6 P3 F2 J1 - 10 J20 12 S1 S1
4 DC4 Slot 1-P3 4 P3 E4 J1 - 4 J32 13 L29 L30
8 DC8 Slot 2-P3 4 P3 I4 J1 - 8 J33 14 L27 L28
16 DC16 Slot 2-P3 8 P3 J4 J1 - 16 J45 15 L25 L26
12 DC12 Slot 1-P3 8 P3 F4 J1 - 12 J37 16 L23 L24
J3-PA
17 DC17 Slot 1-P3 9 P3 G1 J2 - 1 J22 1 L22 L22
21 DC21 Slot 2-P3 9 P3 K1 J2 - 5 J38 2 L20 L20
29 DC29 Slot 2-P3 13 P3 L1 J2 - 13 J36 3 L17 L17
25 DC25 Slot 1-P3 13 P3 H1 J2 - 9 J23 4 L14 L14
1 DC1 Slot 1-P3 1 P3 E1 J1 - 1 J24 5 L11 L11
5 DC5 Slot 2-P3 1 P3 I1 J1 - 5 J25 6 L8 L8
13 DC13 Slot 2-P3 5 P3 J1 J1 - 13 J46 7 L5 L5
9 DC9 Slot 1-P3 5 P3 F1 J1 - 9 J39 8 L2 L2
Note:

Dark grey cells in the DC column indicate combined elements. These should supply approximately double the MC bias current.

Figure 3. GEM DOCK Connector