• Object ID: 00000018WIA3005DE20GYZ
  • Topic ID: id_13105938 Version: 4.0
  • Date: Oct 9, 2019 5:31:24 PM

Lower extremity positioner FRUs and replacements

Personnel Requirements

Required Persons Procedure Timing
115 minutes

Overview

This document contains the list of FRUs and replacement parts for the 16-channel Flex Coil Foot/Ankle/Knee Positioner (PN 5437192, M7005BE). It also includes the latch replacement procedure.

The positioner can be used on GEM (flat) tables or curved tables. To use on a GEM table, remove the two curved pieces attached to the base plate of the positioner. See Figure 2.

Figure 1. Latch (Unlocked) on Positioner
Item Description
1Positioner base plate
2Positioner side wall
3Latch (open position)
4Screw holding lever to base (do not remove)
Figure 2. Positioner Base Plate
Item Description
1Positioner base plate. (Remove two screws on each side to modify the positioner for use on a GEM table.)

FRUs and Other Accessories

Table 1. 16-Channel Flex Coil Foot/Ankle/Knee Positioner FRUs
Description GE Part Number

Small Mechanical Part FRU Kit

  • Snap locks (x2)

  • Screw caps

  • Latch assembly (x2)

  • Left and right support assemblies

  • Spare fasteners

5504484

The following consumable parts are not intended to be covered by any GE service contract. If replaced, they should be billed to the customer.

Table 2. Consumable Parts
Description GE Part Number Image
Pad and Straps FRU Kit5504483
Pad and Straps FRU Kit Contents GE Part Number

Foot Strap

5456263

Ankle strap

5453776

Foot support heel pad

5457814

Foot support sole pad

5457033

Heel pad

5455481

Velcro hook pad 1

5503314

White Velcro hook tape #10

5503314-10

White Velcro hook pad 2

5503314-2

White Velcro hook pad 3

5503314-3

White Velcro hook tape #4

5503314-4

White Velcro hook tape #5

5503314-5

White Velcro hook tape #6

5503314-6

White Velcro hook tape #7

5503314-7

White Velcro hook tape #8

5503314-8

White Velcro hook tape #9

5503314-9

White Velcro loop tape 1

5503315

White Velcro loop tape #2

5503315-2

Anti-skid pad 2

5462302-2

Foot-ankle positioner anti-skid pad

5462302

Anti-skid pad 3

5462302-3
Foot and Coil Holder FRU Kit
  • 90 degree foot holder

  • U-channel coil holder

  • 45 degree forefoot holder

  • Snap locks (x2)

5504485

Consumable

Loctite 242 Threadlocker (GE P/N 46-170686P1)

Latch Replacement

The latch replacement assembly is part of the small mechanical part FRU kit.

  1. The base plate of the positioner has two covers that protect the slot and base of each latch. Using a flat-blade screwdriver, remove the plastic screws that attach the slot cover to the base plate of the positioner.

  2. Remove the old latch by removing the brass screw on the bottom of the latch. The brass screw was seated with Loctite, so you must turn the screwdriver firmly to break the adhesive.
    Figure 3. Remove Brass Screw to Remove Latch
    ItemDescription
    1Positioner base plate
    2Stabilizer foot of latch attached with brass screw
  3. Install the new latch with the latch in the unlocked position. Latch the stabilizer foot to lay in the channel as shown in Figure 3.

    1. Open the latch lever to the unlocked position (oriented 180 degrees to the lock position).

    2. The latch assembly sits on top of the curved washer, with the bottom of the latch assembly in the curved portion of the washer. Hold the latch assembly on top of the base plate.

    3. Position the latch stabilizer foot in the channel on the bottom of the positioner plate and line it up with the latch assembly.

    4. Apply Loctite 242 to the brass screw and install it through latch stabilizer foot on the base plate.

    5. Torque the brass screw to 5 inch-pounds (or snug, plus ¼ turn).

  4. Reinstall the base plate cover.