Replacing the rocker clamp, rod link connector, or shaft rocker
Removes and replaces the defective rocker clamp, rod link connector, or shaft rocker on the patient table.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Link Rod Connector Kit | 1 | 5769330-2 (GEM) 5769330 (Non-GEM) | - |
| Shaft Rocker | 2 | 5147700-2 (GEM) 5147700 (Non-GEM) | - |
| Rocker Clamp | 2 | 5392170 (GEM) 5147701 (Non-GEM) | - |
| Warning | |
|---|---|
About this task
-
The rod link connector acts like a mechanical fuse. If the rod is bent, investigate and correct the problem. If the root cause is not corrected, the rods will continue to bend.
- It is recommended to check the alignment of the bosses before checking for other potential causes. See Boss alignment for more information.
-
Only straighten the rods to keep customer site working. Order replacement rods quickly and install the rods soon after delivery. Even the slightest bend in the rod will weaken the rod’s strength.
-
Do not try to readjust the length of the rod. Rod length for both DV and GEM systems have now been set to the same length.
-
New rods must be installed. The old rods cannot be straightened for reuse. The adjustments made will not deter the rods from bending again.
-
Always replace both rods at the same time.