Removing the Digitized Per Pin (DPP) Receiver Distributor (Rx Dist)
Remove the Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) from the magnet outer electronics.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
15 minutes
-
Tools and test equipment
Item
Quantity
Part number
Manufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set
1 Kit
5113258
-
Nonmagnetic Titanium Service Tool Kit, Large Set
5112581
-
Procedure
CAUTION
Strong magnetic field
The Digitized Per Pin (DPP) Receiver Distributor (Rx Dist) has no ferrous material and will not be attracted to the magnet, but the eddy currents can cause resistance when you pull the DPP Rx Dist away from the magnet.
When removing and replacing the DPP Rx Dist, keep the receiver as far away from the magnet as possible while moving it.
Loosen four screws and remove the DPP Rx Dist.
Note: The screw holes for attaching the DPP Rx Dist are a keyhole shape or slotted so it is not necessary to remove the four screws. Just loosen them.