- Optima MR450w BASE 1.5T System Service Methods
- 5690012-2EN Revision 3
- 00000018WIA308C9F20GYZ
- id_131064115.1
- Jun 8, 2021 3:49:54 PM
Replacing the Digitized Per Pin Remote Receiver (DPP RRx)
Removal and installation instructions for replacing the Digitized Per Pin Remote Receiver (DPP RRx) in the magnet outer electronics.
Prerequisites
The following tables represent a comprehensive roll-up of the personnel, time, tools, test equipment, and replacement parts you need to complete all possible tasks in this procedure.
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Digital Voltmeter (DVM) | 1 | 46-194427P284 | - |
| Red Warning LOTO Tag | 2 | 46-194427P322 | - |
| Multi-Locking Device (if multiple service personnel are involved) | 1 | 46-194427P313 | - |
| Red LOTO Personal Lock Wrap | 2 | 2393068 | - |
| Brass Master Padlock with Brass Shackle | 2 | 46-194427P320 | - |
| MCRv Tool Kit 3.0T | 1 | 5182417-4 | - |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 Kit | 5113258 | - |
| Nonmagnetic Titanium Service Tool Kit, Large Set | 5112581 | - | |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 1 | - | ||
| Safety | ||||||||
|---|---|---|---|---|---|---|---|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. | ||||||||
| ||||||||