Lowering the raised dockable table bellows

Lowers a previously raised dockable table bellows cover and attaches the bellows to the table base covers.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Release the Velcro strips holding the bellows cover in the raised position.
  2. Lower the bellows cover.
  3. Secure the bellows cover to the table base side covers with M6 x 20 Button Head Collar Screws.