• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA3068E030GYZ
  • Topic ID: id_12374633 Version: 1.27
  • Date: Oct 11, 2021 3:47:43 PM

Replacing the hydraulic rotary pump

Replaces the hydraulic rotary pump of the dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 to 105 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
11/16 inch Open-End Wrench2--
7/8 inch Open-End Wrench1--
Needle-Nose Pliers1--
Slotted Screwdriver1--
MCRv Tool Kit 3.0T15182417-2-
MCRv Tool Kit 1.5T15182417-
Consumables
ItemQuantityPart numberManufacturer
Loctite #24250 cc bottle--
Absorbent TowelsAs needed--
Replacement parts
ItemQuantityPart NumberManufacturer
Patient Table Hydraulic Rotary Pump Assembly15341568-
Patient Table Hydraulic Rotary Pump Female Coupler15336201-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.