Digital volt meter
(DVM) with alligator clip leads
1
-
-
-
Non-ferrous tool
kit
1
-
-
Replacement parts
Item
Quantity
Effectivity
Part number
Manufacturer
Gen1 Gradient Filter Assembly (RoHS)
1
-
5166619-50
-
Table 1. Safety
DANGER
FATAL ELECTRIC SHOCK HAZARD!
THE GRADIENT Amplifiers ARE CONSTANT CURRENT SOURCES AND
WILL SEND maximum CURRENT TO ANY LOAD including any person touching
the gradient filter connectors.
TO PREVENT FATAL ELECTRIC SHOCK, ENSURE THAT POWER IS OFF
TO the gradient subsystem before CONTINUING WITH THIS PROCEDURE.
Warning
Potential Projectile Hazard!
ferrous materials can become dangerous projectiles in the
presence of the magnetic field produced by the magnet.
Do not bring any ferromagnetic tools or equipment into
the magnet room.
About this task
Overview
The gradient filter attaches to the secondary pen wall (SPW)
from the equipment room. The thumbscrews on the filter attach to PEM
nuts on the SPW.
Topic ID: id_SL2018500-1062451
Removing GEN1 Gradient Filter
Procedure
DANGER
FATAL ELECTRIC SHOCK HAZARD!
THE GRADIENT Amplifiers ARE CONSTANT CURRENT SOURCES AND WILL SEND maximum CURRENT TO ANY LOAD, including any person touching the gradient filter connectors.
TO PREVENT FATAL ELECTRIC SHOCK, ENSURE THAT POWER IS OFF TO the gradient subsystem before CONTINUING WITH THIS PROCEDURE.
Perform LOTO on the PGR cabinet. Refer to the Applying LOTO - ISC procedure.
Warning
Potential Projectile Hazard!
ferrous materials can become dangerous projectiles in the
presence of the magnetic field produced by the magnet.
Do not bring any ferromagnetic tools or equipment into
the magnet room.
From the scan room side of the gradient filter, remove the plastic
gradient filter terminal covers.
Figure 1. Gradient Filter Assembly Gen1
Note:
Two wrench technique: Use two wrenches
when removing the cables from the gradient input/output connectors
on each side of the SPW. One wrench prevents the post from turning,
while the other loosens the nut. If you try to loosen the nut with
only one wrench, the post itself could turn and damage the gradient
filter.
Remove the cables attached to the terminals in the scan room.
From the equipment room side of the gradient filter, remove
the plastic gradient filter terminal covers and remove the terminal
cables. Use the two wrench technique to remove
the cables.
While in the equipment room at the SPW, remove the screws securing
the gradient filter assembly to the SPW panel.
Remove the gradient filter and set aside.
Figure 2. Gradient Filter Thumbscrews
Topic ID: id_SL2018510-1062451
Installing GEN1 Gradient Filter
Procedure
Lift the gradient filter onto the guide pins on the SPW and tighten the thumb screws. When done, use a screwdriver to tighten.
Attach the cables to the gradient filter connections on both sides of the filter. Start with the GND, Z-axis, Y-axis, and X-axis cables and ground wires. Snugly tighten each cap nut to attach the correct axis cable to the correct gradient filter connection. The lug connections between axis and polarities should not be touching.
Figure 3. Gradient Terminals
Note:
Red cable indicates positive (+) wires, black cables indicate negative (-) wires.
Two wrench technique: Use two wrenches when attaching the cables from the gradient input/output connectors on each side of the SPW. One wrench prevents the post from turning, while the other tightens the nut. If you try to tighten the nut with only one wrench, the post itself could turn and damage the gradient filter.
Note:
Make sure to tighten each cable as you attach it, because you cannot reach all cables after attaching them.
Tighten the cap nuts and verify that nuts are not loose.
Verify that all connections are tight, correct, and not touching each other. Make sure each screw is fixed, each nut is tight, and each cable is not detached from terminal.
Figure 4. Cables Connected
Replace the plastic gradient filter terminal covers.
Remove LOTO from the PGR cabinet. Refer to the Removing LOTO - ISC procedure.