Replacing the hydraulic filter
Removal and installation instructions for replacing the hydraulic filter in the dockable patient table.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 40 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 3/4 inch Open-End Wrench | 2 | - | - |
| Ear Syringe | 1 | - | - |
| Standard Tool Kit with Phillips Screwdrivers | 1 | - | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Hydraulic Oil | 1 | See FRU Manual | - |
| Towels | As needed | - | - |
| Loctite® 569 | 1 bottle (50 ml) | 46-230614P1 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Hydraulic Oil Filter And Gasket | 1 | See FRU Manual | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Overview
Do the following procedure as directed to maintain smooth, responsive operation of the patient table hydraulics.