Doing a check scan
A check scan is done at the conclusion of many procedures. This functional check assures the unit is in good clinical working order.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 15 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| DQA III Phantom | 1 | 2321556 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
Make sure that you use the DQA phantom, so that the image can be analyzed for correct orientation.
| Notice | |
|---|---|