Calibrating High Order Eddy Current
Use HOEC to characterize the time-dependent magnetic field errors in the bore up to at least the third order.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | 10 minutes |
| Tools and test equipment | ||||
|---|---|---|---|---|
| Item | Quantity | Part number | Manufacturer | |
| LVShim Phantom Assembly | 1 | - |
2298119 | - |
| Shim power supply | 1 | - |
46-260777G3 5395802 | - |
| Shim cable kit | 1 | - |
2135558 | - |
| LVShim Phantom Inner Sphere | 1 | 2274476 | - | - |
| Universal Phantom Pad | 1 | 46-328383P1 | - | - |
| Required conditions |
|---|
| Confirm that LVShim-Gradshim calibration is complete. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
| Warning | |
|---|---|