Test
|
Procedure
|
Criteria for Passing
|
Notes
|
| PM Schedule | Run once at time of system turnover. |
|
|
| Updates | Go to CSD, PM section and execute PM Patch.
Document can be found in PM Assist | Broadband site: Informs FE of available Upgrades
needed, and Upgrades that have been automatically downloaded, then
prompts FE to load Upgrades. | No Customer notification |
| LVShim | Go to CSD, PM section and execute LVShim. Document
can be found in PM Assist | Must meet Customer Acceptance values in following
links (BRM WB (Whole Body)) | Customer receives warning message for failures
21 days following PM. IBIS logs failure |
| SPT | Go to CSD, PM section and execute SPT. Document
can be found in PM Assist |
There can be no Severe failures in the following tests:
Head SNR Head Coherent Noise Head FSE Stability Head GRE Stability Head Eddy Currents Body Shimming Body SNR Body FSE Stability Body GRE Stability
| Customer receives warning message for failures
21 days following PM. IBIS logs failure |
| White Pixel | Go to CSD, PM section and execute EPT. Document
can be found in PM Assist | Run test only. White Pixel failures not counted
by IBIS. This is for trending/data-collection purposes only. | No Customer notification |
| PM Check | Run this script after all checks noted above
have been executed |
|
|