Testing MRU finalization
Procedure
- Make sure that any cable removed during testing has been re-connected and close MRU.
- Make sure there is continuity between the MRU and magnet by doing a heater test:
- Press the Test Heater LED on the MRU. Make sure the Heater Test LED Illuminates.note: Rapid movement or toggling of the Test Heater LED between the A and B position can cause the Heater Actuated LED to turn on.
- Move the test switch to position A and make sure the Test Heater LED illuminates.
- Move the test switch to position B and make sure the Test Heater LED illuminates.
- Press the Test Heater LED on the MRU. Make sure the Heater Test LED Illuminates.